CellOx 325

Copyright

©

 2017 Xylem Analytics Germany GmbH

Printed in Germany.

Содержание WTW CellOx 325

Страница 1: ...OPERATING MANUAL ba25301e06 12 2017 CellOx 325 DISSOLVED OXYGEN SENSOR OPERATING MANUAL...

Страница 2: ...2 ba25301e06 12 2017 CellOx 325 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...4 2 Measuring 7 4 3 Storage 7 5 Maintenance cleaning replacement 8 5 1 General maintenance notes 8 5 2 Outside cleaning 8 5 3 Changing the electrolyte and membrane cap 9 5 4 Cleaning the electrodes 1...

Страница 4: ...Inhaltsverzeichnis CellOx 325 4 ba25301e06 12 2017...

Страница 5: ...n rivers lakes and wastewater Applications in water laboratories BOD measurements 1 WP 90 membrane cap filled with electrolyte solution 2 Shaft 3 Connection head 4 Gold working electrode cathode 5 Ins...

Страница 6: ...order to avoid slight injuries or damage to the instrument or the environment 3 Commissioning Scope of delivery D O sensor CellOx 325 filled with electrolyte solution and ready to operate Calibration...

Страница 7: ...water pipe stream Slowly pull the sensor through the water by hand lake container or Use a flow aid such as a magnetic stirrer with stirring device see chapter 8 WEAR PARTS AND ACCESSORIES 4 3 Storage...

Страница 8: ...tact with the skin rinse thoroughly with water and immediately change contaminated clothing If it comes into contact with the eyes rinse thoroughly with water and consult a doctor Note the safety data...

Страница 9: ...is damaged the electrolyte solution is exhausted in case of a leak message by the measuring instrument Changing the electrolyte and membrane cap Unscrew the membrane cap Caution Electrolyte solution F...

Страница 10: ...membrane head fingertight using a paper towel Excess electrolyte solution is forced out of the ventilation area Check the filling Inspect the face surface No air bubbles may be present within the dash...

Страница 11: ...ane cap Caution Electrolyte solution For disposal of the membrane head and electrolyte solution see section 5 6 Rinse the sensor head with deionized water Using the rough side of the wet SF 300 polish...

Страница 12: ...r for at least 10 minutes Carefully shake off the drops of water Fill a new membrane cap with ELY G electrolyte solution Remove any air bubbles by carefully tapping the membrane head Additionally you...

Страница 13: ...he sensor in the sodium sulfite solution for more than 2 minutes Danger of sensor poisoning Test criterion The sensor is OK if the measuring instrument displays 1 D O saturation after 2 minutes Hold t...

Страница 14: ...l according to the safety data sheet 6 What to do if Error symptom Cause Remedy The sensor is in the air and the display shows 0 0 mg l or 0 O2 No connection between measuring instrument and sensor No...

Страница 15: ...do if 15 ba25301e06 12 2017 Meter displays OFL display range exceeded Short circuit between working electrode and counter electrode Clean the electrodes see section 5 4 If the meter still displays OFL...

Страница 16: ...electrode Lead Shaft Connection head Cable screw joint Membrane head POM Membrane FEP 13 m Sensor head Epoxy PEEK Thermistor housing VA steel 1 4571 Seals FPM Viton Connection cable Lengths 1 5 20 m D...

Страница 17: ...Storage conditions Recommended storing method In the calibration and storage vessel OxiCal SL moist Storage temperature 0 50 C 32 122 F Characteristics when delivered Zero signal 0 1 of the saturatio...

Страница 18: ...ZBK 325 202 706 Accessories Description Model Order no Calibration and storage vessel OxiCal SL 205 362 Magnetic stirrer Oxi Stirrer 300 203 810 Stirring accessory provides a constant defined flow to...

Страница 19: ......