91
Fig. 59
Max. Titration Time
The maximum titration time is used with samples that generate an increased drift in the end and when it is not
possible to achieve stable end values. The max. titration time can be adjusted from 0 to 9999 s. The standard
value is pre-set to 600 s.
Fig. 60
Min. Titration Time
After the minimum titration time has passed, the adjusted stop criteria are checked. The min. titration time can be
adjusted from 1 to 1800 seconds. The standard value is pre- set to 60 seconds. Particularly when worked with a
KF oven the minimum titration time must be higher.
Fig. 61
Stop delay time
Is the time in seconds in which the drift-stop criteria are being checked. The stop delay time can be set between
0 and 60 seconds. The pre-set standard value is 5 s.