94
If the drift change is within the stop drift tolerance during the entire follow-up time, the measurement will be
ended.
Diagram 2:
The criterion stop drift as difference to the start drift, the stop drift tolerance as stability criterion for
the drift.
Only one of the two end criteria stop drift and stop drift tolerance must be met to end the measurement.
4.6.3
Sample identification
In the manual titration and in the preparation of solutions it is possible to input a sample identification. The
possible input includes manual, automatic or no sample description at all.
Fig. 66
For a sample description of the ’manual’, a prompt for the sample description will always be displayed at the start
of the method (Cp. also chapter 3.6, Main menu). For an ‘automatic’ sample description there will be selected a
master description (in the current case this is water, cp. Fig. 67), which will then automatically be numbered
starting on 01.