2 Product description
9
The flow unit CHEM
FLOW
is integrated into the process pipe and
houses a unit in which the sensor is built in.
The sensor is connected to a transmitter and can thus transfer its
measuring results.
The transmitter can be connected with a process control system.
Fig. 2 Process flow
The pressure and the temperature conditions of the process are
relevant for choosing the suitable flow unit. Depending on the
temperature, the flow unit of stainless steel can be used up to a
pressure of 16 bar, the flow unit of plastic can be used up to a
pressure of 6 bar. The process temperature must lie between -10°
and 140°C.
Pay attention to the pressure and the temperature diagrams!
Principally, the unit may be installed in any position. In order to
get reliable measuring results, the characteristics of the chosen
sensor are relevant.
Unit / sensor
Transmitter
Process control
system
Pressure
Temperature
!!!
Installation
position