August 2020 – TD 316 OPERATING MANUAL – Motus Wave Sensor 5729
Page 89
Table 6-1: Passkey protection
6.5
Save and Reset
When the required properties are set, you should send a
Save
command to make sure that the new configuration is
saved internally in the flash memory. The sensor always reads the configuration from the internal flash memory after
reset and power up. The
Save
command takes about 20 seconds to complete (indicated with the character ‘#’).
Always send a
Reset
command when a new configuration has been saved (or switch the power OFF and then back
ON), or else calculated parameters may be corrupted. This forces the sensor to start up with the new configuration
input. If the
Enable Sleep
property is set to
Yes
and the
Comm TimeOut
property is not set to
Always On
the sensor
enters sleep mode after reset.
At startup/reset the sensor performs measurements according to the interval setting if the mode is
Smart Sensor
Terminal.
If
Enable Text
is set to
Yes
,
the
Startup Info
is presented.
If the
Save
command is executed the new setting will be stored in the internal Flash memory.
Property changes will be lost when the sensor is reset or loses power unless you type the
Save
command.
Refer to Figure 6-2. The number of parameters in the list depends on which parameters are enabled.
Figure 6-2: Save and reset in Tera Term