4

 

MVN Analyze as Master and as Noraxon TeleMyo slave 

4.1

 

Start and Stop a recording 

4.1.1

 

Hardware connections 

 

 

In addition to the normal MTw hardware setup, connect the BNC connector from 
the TeleMyo DTS to Sync OUT 1 on the Awinda Station. 
 

 

Connect the USB connection between the TeleMyo DTS and the PC. 

 

 

When switched on, the DTS will display “USB ready”.

 

 
 

4.1.2

 

Software Setup Noraxon MyoResearch 

The example given below is for is gait analysis, measuring the medial gastrocnemius, 
tibialis  anterior,  semitendinosus  and  the  rectus  femoris.  For  an  8-Channel  EMG 
system, Channel 9 is selected as the synchronisation line in the Noraxon MyoResearch 
Software. If a 16 channel system is in use, this is the 17th channel. 
Return  to  the  main  menu,  under  >Measuring  Options,  go  to  >Recording  Options, 
then>Triggering tab. 
 

 

Figure 6 - Noraxon "Recording Options" screenshot. 

Check the check box beside “Start Recording”;

 

 

Go to the 

drop down menu beside “When Channel”, select “Sync”;

 

 

Select Rises Above (ensure that this is also the direction indicated on 
the mini-receiver); 

 

Input e.g. 0.5V 

 

Longer than 5ms 

Содержание Awinda Station

Страница 1: ...ion in this document is subject to change without notice Xsens MVN MotionGrid MTi MTi G MTx MTw Awinda and KiC are registered trademarks or trademarks of Xsens Technologies B V and or its parent subsi...

Страница 2: ...xon TeleMyo slave 6 4 1 Start and Stop a recording 6 4 1 1 Hardware connections 6 4 1 2 Software Setup Noraxon MyoResearch 6 4 1 3 Xsens Software 8 5 Noraxon TeleMyo as Master and MVN Analyze as slave...

Страница 3: ...m the Master software to start recording 4 Start the recording on the Master software Then both systems will start recording at the same time 5 Stop recording on the master software Then both systems...

Страница 4: ...re Wireless EMG transmitters 1 TeleMyo DTS plus antenna 1 TeleMyo mini receiver to use Noraxon as Master MVN Awinda or MVN Link Awinda station BNC BNC cables Figure 2 Awinda Station 4 BNC ports SYNC I...

Страница 5: ...can detect polarity changes on the input lines When a trigger is detected on one of the input lines the Awinda or Sync Station can be configured to perform a specific action 3 2 Sync OUT Sync OUT ena...

Страница 6: ...rode or sensor site to a belt worn receiver The TeleMyo 2400R G2 Mini Receiver is an optional accessory item for the portable TeleMyo 2400T G2 Transmitter System It is TTL compatible facilitating sync...

Страница 7: ...is gait analysis measuring the medial gastrocnemius tibialis anterior semitendinosus and the rectus femoris For an 8 Channel EMG system Channel 9 is selected as the synchronisation line in the Noraxon...

Страница 8: ...e The Noraxon software initialises the EMG signals When this screen is reached click record on the bottom left hand side of the screen on the Noraxon software Figure 7 Noraxon system ready to begin re...

Страница 9: ...following settings 1 In the Configuration window in the Sync tab you should select Strat recording out Stop Recording out Out 1 Polarity Rising Edge Skip Factor 0 Skip First 0 Pulse width 10ms Trigge...

Страница 10: ...T port of TeleMyo mini receiver to BNC connection Sync IN 1 of Awinda Station Manual trigger pulse jack connector to Sync IN port of TeleMyo mini receiver Connect the external antenna to the TeleMyo D...

Страница 11: ...y Rising Edge Skip Factor 1 Skip First 0 Delay 0 ms Trigger once uncheck In 1 Polarity Rising Edge Skip Factor 1 Skip First 1 Delay 0ms Trigger once uncheck To initialize recording click the record bu...

Страница 12: ...te of 120Hz you will realise that the maximum frame rate that the output signal will have is 120Hz Practically this entails that every other frame of Xsens will be synchronize with the frame of the ot...

Отзывы: