Series EJ, ET, EY, FJ & FR
102002-177 Rev M2
29
\1020\02177M2.doc 17 Jun 2019
PROGRAM REM/LOC Pushbutton:
Pressing this button toggles the voltage and current programs between local and
remote. It functions either in standby or when the HV is ON.
Remote program is analog by default. When the RS232, USB, or optional Ethernet
connector is connected to a computer running control software, operation will
automatically switch from analog to digital when in remote mode. The USB will
take priority over the RS232 when the USB connector is connected, and the Ethernet
(optional) will take priority over both the RS232 and the USB when connected.
When in local mode the monitors and status can still be read over any digital
connection.
Upon turning the AC power switch ON the unit will default to the previous setting of
the switch when the unit was last powered.
The NC option requires continuous remote operation. However the digital/analog
remote priorities remain the same.
PROGRAM REM/LOC Indicator:
Green LED indicator that is lit whenever the unit is in remote program mode.
ENABLE REM/LOC Pushbutton:
Pressing this button toggles the HV enable between Local and Remote. It functions
either in standby or when the HV is ON.
The HV Enable is switched from always on (local) to external analog or digital. The
remote HV Enable is analog by default and requires an active pull-up to generate HV
output. When the RS232, USB, or optional Ethernet connector is connected,
operation will automatically switch from analog to digital when in remote mode.
The USB will take priority over the RS232 when the USB connecter is connected
and the Ethernet (optional) will take priority over both the RS232 and the USB when
connected.
Upon turning the AC power switch ON the unit will default to the previous setting of
this switch when the unit was last powered.
The NC option requires continuous remote operation. However the digital/analog
priorities remain the same.
ENABLE REM/LOC Indicator:
Green LED indicator that is lit whenever the unit is in remote enable mode.