Series EJ, ET, EY, FJ & FR
102002-177 Rev M2
19
\1020\02177M2.doc 17 Jun 2019
15.
Set external I-PROGRAM pot to a level that is greater than the amount that
the connected load will require (any setting above zero if no load is
connected). Note: A setting above zero is required for HV generation even if
no load is connected.
16.
Depress HV ON pushbutton. The HV ON indicator should illuminate. (“NC”
option users: HV is generated upon AC power on)
17.
Rotate external V-PROGRAM pot until voltage meter indicates desired
output voltage.
18.
To shut the HV OFF, press the STANDBY switch. The HV on indicator will
extinguish. (Not for “NC” option units)
FOR REMOTE DIGITAL CONTROL:
19.
Make sure the PROGRAM and ENABLE REM/LOC switches are set to
remote. The REM/LOC indicators should be illuminated. (Not applicable for
“NC” option)
20.
Launch the supplied software on the connected computer.
21.
Depress HV ON pushbutton. The HV ON indicator should illuminate. (“NC”
option users: HV is generated upon AC power on)
22.
Using the supplied software, set voltage and current programs to desired
levels. Select HV ON, then Send Program.
23.
To shut the HV OFF, press the STANDBY switch. The HV on indicator will
extinguish. (Not for “NC” option units)
Refer to page 33 for complete REMOTE DIGITAL CONTOL operating
instructions.
24.
To shut down supply, set POWER SWITCH to the off (“0”) position.
WARNING!
DO NOT HANDLE THE LOAD OR EXPOSED HIGH VOLTAGE
TERMINATIONS OR ATTEMPT TO MAKE OR REMOVE ANY
CONNECTIONS TO THE SUPPLY UNTIL THE LOAD AND/OR SUPPLY
HAS BEEN DISCHARGED (GROUNDED). AN UNLOADED SUPPLY MAY
TAKE UP TO 15 SECONDS TO FULLY DISCHARGE.