XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
66
The Mixing cup is detected
as empty when filled
Liquid sensor pins are misaligned or
defective S11B
is assumed when no cleaning fluid is
detected in mixing cup.
Check that the pins do not touch the
edges of the mixing cup or are short-
circuited.
Consult the chapter
“Liquid Sensor PCB”,
AP-869" on page 56
Replace “Liquid Sensor PCB” AP-
869
Measure VCC 5V between Pin 1
(GDD) and Pin 4 (VCC)
The pump V35 is defective.
Check if the pump V35 connections
are properly inserted (CA-008 cable).
Check if the Cable CA-008 is not
damaged resulting to a short-circuit
Measure the voltage at the coil
connectors 24VDC.
Measure the coil resistance (≈100Ω).
Change if defective (AO-346)
Disassemble and clean the valve at
the piston.
Change the plunger seat gasket
(MR-150).
The suction cleaning liquid
is detected as "low" while
the bottle is full (Yellow
bottle)
The photo sensor S15 measuring the
suction clean fluid in the dispenser is
not properly aligned, obturated or is
defective.
Measures the content of the suction
clean dispenser located in the
service panel
Check that there is no gap between
the sensor and the dispenser.
If necessary, adjust the liquid sensor
and secure that the dispenser is
properly installed.
Verify that the rubber support is
good condition and does not
obturate the 2 photos electric
sensors. Replace the rubber support
if damage
Replace the liquid sensor AN-374
Check that the CA-008 cable is not
defective or damaged.
Changed the cable if defective
Unpleasant smells from the
suction
The suction cleaning liquid dispenser
is empty and the “check Yellow
bottle” warnings are ignored.
Insert a new dispenser
The liquid sensor pins are misaligned
or defective S11B
Check that the pins do not touch the
edges of the mixing cup or are short-
circuited.
Consult the chapter
“Liquid Sensor PCB, AP-869” on
page 56
Measure the level of the cleaning
liquid on the dispenser.
To check that the liquid is extracted
from the cartridge during the suction
cleaning process, make a small mark
on the cartridge at the current liquid
level. After performing a suction
cleaning session, check that the
liquid level has dropped during the
process. Otherwise, check the
reasons described down below.
Содержание 4
Страница 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Страница 24: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 24 BRIDGE PCB CONNECTORS J2 TO J36 Point of measurement...
Страница 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Страница 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Страница 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Страница 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Страница 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...