XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
75
Chair synchronisation:
It is mandatory to synchronize the chair.
•
After replacing the AN-368 PCB situated in XO4 & XO FLEX bridge.
•
After replacing the lithium battery.
•
After a firmware update.
To synchronise the chair, proceed as follows
:
1
Power the unit on and wait until it is ready. Continue with step 2 or see alternative procedure.
2
Press C (configuration switch) twice. Display shows
“
LIGHT
”
3
Press foot control joystick west (left). Display shows
“
UNIT
”
4
Press foot control joystick south (against yourself). Display shows
“
CHAIR
”
5
Press foot control joystick east (right). Display shows
“
RINSE=0
”
6
Press foot control joystick south. Display shows
“
SYNC
”
7
Press foot control joystick east. Display shows
“
NO
”
8
Press foot control pedal down. Display shows
“
YES
”
9
Press foot control joystick west. Display shows
“
SYNC
”
and the chair moves.
10
Press the configuration switch. Display shows same info as in 1, and calibration has ended when
the chair stops.
Note that this description refers to firmware version 1.51 or later.
Alternative procedure:
Same as above, but after step 1 move the chair (even slightly) and execute step 2 quickly after.
Display shows “SYNC”. Continue from step 7.
Содержание 4
Страница 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Страница 24: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 24 BRIDGE PCB CONNECTORS J2 TO J36 Point of measurement...
Страница 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Страница 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Страница 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Страница 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Страница 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...