6
6-7
Understanding calibration
6.
Configure the densitometer to factory defaults and save the settings. (See the
densitometer manual for details.)
Factory default settings for the X-Rite DTP32 are:
7.
Turn on the PowerBook.
8.
In the PowerBook Setup Control Panel, select Normal for Modem Controls.
9.
Calibrate the densitometer, as described below.
T
O
CALIBRATE
THE
X-R
ITE
DTP32:
1.
On the X-Rite DTP32, simultaneously press the two buttons marked MENU.
The words
MAIN MENU
appear in the display.
2.
Press the p1 key until p2 appears.
3.
Press the cal key.
CALIBRATION
appears in the display.
4.
Press the den key.
CALIBRATING TRANSMISSION
is displayed, followed by the words
INSERT CAL STRIP
.
5.
Position the strip guide at 5.
6.
Insert the arrow end of the Auto-Cal strip into the entrance of the X-Rite DTP32 until
the roller starts pulling the strip.
Page 1
LANG (ENG)
TONE (loud)
TYPE (ser)
Page 2
BAUD (9600)
HAND (xon)
AXMT (on)
Page 3
DPT (on)
SEP (spc)
DLIM (crlf )
Page 4
DEF (off )
X10 (off )
DAP (off )
Page 5
M/M (max)
LOCK (off )
Strip guide
Содержание Digital copier printers
Страница 1: ...US ER G UI D E for Xerox digital copier printers Fiery SI Color Server...
Страница 43: ...33 12 Printing with PC Compatible Computers 7 Click the Graphics tab...
Страница 51: ...44 3 Scanning documents 6 Click Prescan The image appears on the left side of the window...
Страница 104: ...66 26 Color Calibration The rest of the curve adjusts to match the endpoint...
Страница 105: ...66 27 Using the Fiery XJ Print Calibrator TO EDIT CUSTOM CURVES 1 Click Custom to edit the curve...