Recommended separation distances between

portable and mobile RF communications equipment and the STIMPOD NMS460

The STIMPOD NMS460 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the STIMPOD NMS460  can 
help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the STIMPOD NMS460 as 
recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power

W

Separation distance according to frequency of transmitter 

m

150 kHz to 80 MHz

Not applicable

80 MHz to 800 MHz 

d = 1,2    

800 MHz to 2,5 GHz 

d = 2,3   

0,01 

-

0,12

0,23

0,1 

-

0,38

0,73

-

1,2

2,3

10 

-

3,8

7,3

100 

-

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency 
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1  

At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2  

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Guidance and manufacturers declaration – electromagnetic immunity – for equipment and systems that are non- 

life supporting

The STIMPOD NMS460  is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS460  should ensure that it is used in such 
an environment

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Radiated immunity
80MHz - 2.5GHz

80MHz – 1GHz @ 3V/m & 
10V/m
1GHz – 2.5GHz @ 10V/m

80MHz – 1GHz @ 3V/m & 
10V/m
1GHz – 2.5GHz @ 10V/m

Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT and should be used no closer to any part of the 
equipment, including cables, than the recommended separation distance.

vi

Содержание STIMPOD MMS460

Страница 1: ...PRODUCT CODE XT 46006 EN XM400 31A04 07...

Страница 2: ...i...

Страница 3: ...be observed to verify normal operation in the configuration in which it will be used The patient should avoid contact with metallic objects that are grounded produce an electrical conductive connectio...

Страница 4: ...e stimulators are a symptomatic treatment and as such suppresses the sensation of pain that would otherwise serve as a protective mechanism Effectiveness is highly dependent upon patient selection by...

Страница 5: ...healthcare pro fessional only This equipment system may cause radio interference or disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or re...

Страница 6: ...e d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separatio...

Страница 7: ...pplicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separa...

Страница 8: ...er 4 1 8 Shutdown 4 1 9 Symbols on device battery clip 4 2 Setting up Device Defaults 5 2 1 Languages 5 2 2 Stimulation Mode 5 2 3 Pulse Width Options 6 2 4 Stimulating Frequency 6 2 5 Countdown Timer...

Страница 9: ...n those supplied with the STIMPOD NMS460 may result in serious injury NOTE Standard ECG Electrodes can be used for the anode Such ECG electrodes need to conform to ANSI AAMI EC12 on disposable ECG ele...

Страница 10: ...s to Enter in setup menu Stimulating LED indicator Flashing Green Stimulus delivered Flashing Red Open Circuit detected Menu Pulse Width Button Press to toggle between Pulse Widths Press and hold to a...

Страница 11: ...too high higher than 13 kOhm This is most likely caused by a poor grounding electrode skin connection Pause This warning informs the user that the PAUSE button was pressed The STIMPOD NMS460 will pau...

Страница 12: ...ircuit Detection A few milliseconds prior to the actual stimulation an impedance measurement is made to detect whether the connection between the STIMPOD NMS460 and the patient comprises a closed circ...

Страница 13: ...croll down the menu to EXIT MENU and press enter A conformation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you to the previous...

Страница 14: ...ter A confirmation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you back to the previous screen 2 5 Countdown Timer DEFAULT 10 mi...

Страница 15: ...rstheusertheoptiontoenteruserinformation Twolinesof20 characters each may be entered This user information will be displayed for two seconds when unit is switched on as per example Use the Wheel to na...

Страница 16: ...ld flash GREEN and if sound is enabled in the menu a beep should be heard each time a stimulus is delivered Stimulus should occur at the set frequency 1 2 or 5 Hz Use the adjusting wheel and slowly in...

Страница 17: ...er Consumption 17mA Weight 130g Dimensions 145mm x 90mm x 30mm Operating Temperature 10 40 Celsius Storage and Transport Temperature 0 50 Celsius Operating Humidity 90 Relative Humidity 3 3 Cleaning a...

Страница 18: ...e NMS 460 PERIPHERAL NERVE STIMULATOR ONLY XT 46000 No cables or accessories NMS 460 PERIPHERAL NERVE STIMULATOR KIT XT 46001 Including Mapping Probe Carry Case IFU NERVE MAPPING CABLE XT 46003 CARRY...

Страница 19: ...11...

Страница 20: ...Unit 102 The Tannery Industrial Park 309 Derdepoort Rd Silverton Pretoria South Africa 0184 Tel 27 0 12 743 5959 Fax 27 0 86 547 0026 E mail support xavant com Web www xavant com...

Отзывы: