Xavant STIMPOD MMS460 Скачать руководство пользователя страница 3

Manufacturer

Xavant Technology Pty (Ltd)

 

Unit 102, The Tannery Industrial Park, 309 Derdepoort Rd
Silverton, Pretoria, South Africa, 0184
Tel:  

+27 (0) 12 743 5959 

Fax:  

+27 (0) 86 547 0026

E-mail: [email protected]
Web: www.xavant.com

Legal Representative in the EU

Emergo Europe

 

Prinsessegracht 20
2514 AP The Hague
The Netherlands
Tel: 

+31 70 345 8570

Fax: 

+31 70 346 7299

Indications for use:

The STIMPOD NMS460 is a Transcutaneous Electrical Nerve Stimulation device 
used for symptomatic relief and management of chronic intractable pain and/
or as an adjunctive treatment in the management of post-surgical pain, post 
traumatic acute pain problems, as well as an adjunct for pain control due to  
rehabilitation.

Contraindications:

•  Known neurological disorders. 
•  Do not use this device on patients who have a cardiac pacemaker, implanted de-

fibrillator, or other implanted metallic or electronic device, because this may cause 
electric shock, burns, electrical interference, or death.

•  Do not use this device on patients whose pain syndromes are undiagnosed.

Warnings:

•  Read the entire User Manual before attempting to use the device.  
•  Use of cables or accessories other than those supplied with the STIMPOD NMS460 

may result in serious injury.

•  Maintenance on this device may only be performed by the manufacturer or 

persons explicitly authorized by the manufacturer.

•  Do not use the STIMPOD NMS460 in close proximity to equipment that produces 

strong electromagnetic fields, such as high frequency surgical equipment. The 
cable leads could act as antennae and dangerous currents could be induced as 
a result.

•  The device should not be used adjacent to or stacked with other equipment and 

that if adjacent or stacked use is necessary, the device should be observed to verify 
normal operation in the configuration in which it will be used. 

•  The patient should avoid contact with metallic objects that are grounded, produce 

an electrical conductive connection with other equipment and/or enable capacitive 
coupling.

•  The cables should be positioned in such a way that they do not contact either the 

patient or other cables.

•  Simultaneous connection of a patient to high frequency surgical ME equipment 

and the STIMPOD NMS460 may result in burns and possible damage to the 
stimulator.

•  Operation in close proximity (e.g. 1m) to a shortwave or microwave therapy ME 

equipment may produce instability in the stimulator output.

•  Application of electrodes near the thorax may increase the risk of cardiac fibrillation.
•  If battery acid has leaked into the device essential circuitry may have been 

compromised. In the event of leakage the device must be returned to its 
manufacturer for safety checks and possible repairs. 

•  Do not apply stimulation over the patient’s neck because this could cause severe 

muscle spasms resulting in closure of the airway, difficulty in breathing, or adverse 
effects on heart rhythm or blood pressure.

•  Do not apply stimulation over open wounds or rashes, or over swollen, red, 

infected, or inflamed areas or skin eruptions (e.g., phlebitis, thrombophlebitis, 
varicose veins).

•  Do not apply stimulation over, or in proximity to, cancerous lesions.
•  Do not apply stimulation when the patient is in the bath or shower.
•  Do not apply stimulation while the patient is sleeping.
•  Do not apply the stimulation while the patient is driving, operating machinery, 

or during any activity in which electrical stimulation can put the patient at risk of 
injury.

•  Do not apply stimulation when the leads are open circuited (e.g. when the 

electrodes are held in the user’s hands).

•  Do not remove electrodes when stimulation is being applied (to avoid unintended, 

possibly hazardous, stimulation pathways). 

ii

Содержание STIMPOD MMS460

Страница 1: ...PRODUCT CODE XT 46006 EN XM400 31A04 07...

Страница 2: ...i...

Страница 3: ...be observed to verify normal operation in the configuration in which it will be used The patient should avoid contact with metallic objects that are grounded produce an electrical conductive connectio...

Страница 4: ...e stimulators are a symptomatic treatment and as such suppresses the sensation of pain that would otherwise serve as a protective mechanism Effectiveness is highly dependent upon patient selection by...

Страница 5: ...healthcare pro fessional only This equipment system may cause radio interference or disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or re...

Страница 6: ...e d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separatio...

Страница 7: ...pplicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separa...

Страница 8: ...er 4 1 8 Shutdown 4 1 9 Symbols on device battery clip 4 2 Setting up Device Defaults 5 2 1 Languages 5 2 2 Stimulation Mode 5 2 3 Pulse Width Options 6 2 4 Stimulating Frequency 6 2 5 Countdown Timer...

Страница 9: ...n those supplied with the STIMPOD NMS460 may result in serious injury NOTE Standard ECG Electrodes can be used for the anode Such ECG electrodes need to conform to ANSI AAMI EC12 on disposable ECG ele...

Страница 10: ...s to Enter in setup menu Stimulating LED indicator Flashing Green Stimulus delivered Flashing Red Open Circuit detected Menu Pulse Width Button Press to toggle between Pulse Widths Press and hold to a...

Страница 11: ...too high higher than 13 kOhm This is most likely caused by a poor grounding electrode skin connection Pause This warning informs the user that the PAUSE button was pressed The STIMPOD NMS460 will pau...

Страница 12: ...ircuit Detection A few milliseconds prior to the actual stimulation an impedance measurement is made to detect whether the connection between the STIMPOD NMS460 and the patient comprises a closed circ...

Страница 13: ...croll down the menu to EXIT MENU and press enter A conformation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you to the previous...

Страница 14: ...ter A confirmation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you back to the previous screen 2 5 Countdown Timer DEFAULT 10 mi...

Страница 15: ...rstheusertheoptiontoenteruserinformation Twolinesof20 characters each may be entered This user information will be displayed for two seconds when unit is switched on as per example Use the Wheel to na...

Страница 16: ...ld flash GREEN and if sound is enabled in the menu a beep should be heard each time a stimulus is delivered Stimulus should occur at the set frequency 1 2 or 5 Hz Use the adjusting wheel and slowly in...

Страница 17: ...er Consumption 17mA Weight 130g Dimensions 145mm x 90mm x 30mm Operating Temperature 10 40 Celsius Storage and Transport Temperature 0 50 Celsius Operating Humidity 90 Relative Humidity 3 3 Cleaning a...

Страница 18: ...e NMS 460 PERIPHERAL NERVE STIMULATOR ONLY XT 46000 No cables or accessories NMS 460 PERIPHERAL NERVE STIMULATOR KIT XT 46001 Including Mapping Probe Carry Case IFU NERVE MAPPING CABLE XT 46003 CARRY...

Страница 19: ...11...

Страница 20: ...Unit 102 The Tannery Industrial Park 309 Derdepoort Rd Silverton Pretoria South Africa 0184 Tel 27 0 12 743 5959 Fax 27 0 86 547 0026 E mail support xavant com Web www xavant com...

Отзывы: