Xavant STIMPOD MMS460 Скачать руководство пользователя страница 11

Speaker Volume

Square Wave      Hybrid RF

Battery Status

1.4) 

Screen Layout

  Countdown timer

  Current Setting

Adjust using Wheel.

  Pulse Width Setting

Adjust using Menu / Pulse Width button.

  Warning Screen

  Stimulating Frequency Setting

Adjust using Enter / Hz button.

 Indicators

Insert Cable:

This is the first prompt which the user will encounter as the unit is switched on, and signals that the unit is waiting for the cable to be inserted.

Actual current delivered

The current displayed in this screen is the actual Average Current delivered to the patient. If there is a discrepancy of more than 10% between this 

measured current and the current setting a warning sign “     ” will appear. If the actual current delivered is not the same as the current setting it 

means that the impedance of the circuit between the device and patient is too high (higher than 13 kOhm). This is most likely caused by a poor 

grounding electrode/skin connection.

Pause:

This warning informs the user that the PAUSE button was pressed.

The STIMPOD NMS460 will pause all its activities and wait for the PAUSE button to be pressed again.

Cable not recognised:

This warning informs the user that the inserted cable is not compatible with the STIMPOD NMS460.

Replace Batteries:

This warning informs the user that the batteries are depleted beyond an acceptable level. Continuing to operate the device thus will make it  

unreliable. To prevent this the STIMPOD NMS460 flashes this warning for 4 seconds before switching off.

Open Circuit Detected:

This warning informs the user that the two electrodes (i.e. Ecg electrode and treatment probe) do not form the closed circuit. This warning will be 

accompanied by a red flashing LED, every time the unit attempts to stimulate.

Error:

The STIMPOD NMS460 has detected a component failure. Please send the device back to the manufacturer for repair!

7

3

Содержание STIMPOD MMS460

Страница 1: ...PRODUCT CODE XT 46006 EN XM400 31A04 07...

Страница 2: ...i...

Страница 3: ...be observed to verify normal operation in the configuration in which it will be used The patient should avoid contact with metallic objects that are grounded produce an electrical conductive connectio...

Страница 4: ...e stimulators are a symptomatic treatment and as such suppresses the sensation of pain that would otherwise serve as a protective mechanism Effectiveness is highly dependent upon patient selection by...

Страница 5: ...healthcare pro fessional only This equipment system may cause radio interference or disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as re orienting or re...

Страница 6: ...e d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separatio...

Страница 7: ...pplicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separa...

Страница 8: ...er 4 1 8 Shutdown 4 1 9 Symbols on device battery clip 4 2 Setting up Device Defaults 5 2 1 Languages 5 2 2 Stimulation Mode 5 2 3 Pulse Width Options 6 2 4 Stimulating Frequency 6 2 5 Countdown Timer...

Страница 9: ...n those supplied with the STIMPOD NMS460 may result in serious injury NOTE Standard ECG Electrodes can be used for the anode Such ECG electrodes need to conform to ANSI AAMI EC12 on disposable ECG ele...

Страница 10: ...s to Enter in setup menu Stimulating LED indicator Flashing Green Stimulus delivered Flashing Red Open Circuit detected Menu Pulse Width Button Press to toggle between Pulse Widths Press and hold to a...

Страница 11: ...too high higher than 13 kOhm This is most likely caused by a poor grounding electrode skin connection Pause This warning informs the user that the PAUSE button was pressed The STIMPOD NMS460 will pau...

Страница 12: ...ircuit Detection A few milliseconds prior to the actual stimulation an impedance measurement is made to detect whether the connection between the STIMPOD NMS460 and the patient comprises a closed circ...

Страница 13: ...croll down the menu to EXIT MENU and press enter A conformation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you to the previous...

Страница 14: ...ter A confirmation screen will show your selection with an option to ACCEPT YES will take you back to the main Setup Menu NO will take you back to the previous screen 2 5 Countdown Timer DEFAULT 10 mi...

Страница 15: ...rstheusertheoptiontoenteruserinformation Twolinesof20 characters each may be entered This user information will be displayed for two seconds when unit is switched on as per example Use the Wheel to na...

Страница 16: ...ld flash GREEN and if sound is enabled in the menu a beep should be heard each time a stimulus is delivered Stimulus should occur at the set frequency 1 2 or 5 Hz Use the adjusting wheel and slowly in...

Страница 17: ...er Consumption 17mA Weight 130g Dimensions 145mm x 90mm x 30mm Operating Temperature 10 40 Celsius Storage and Transport Temperature 0 50 Celsius Operating Humidity 90 Relative Humidity 3 3 Cleaning a...

Страница 18: ...e NMS 460 PERIPHERAL NERVE STIMULATOR ONLY XT 46000 No cables or accessories NMS 460 PERIPHERAL NERVE STIMULATOR KIT XT 46001 Including Mapping Probe Carry Case IFU NERVE MAPPING CABLE XT 46003 CARRY...

Страница 19: ...11...

Страница 20: ...Unit 102 The Tannery Industrial Park 309 Derdepoort Rd Silverton Pretoria South Africa 0184 Tel 27 0 12 743 5959 Fax 27 0 86 547 0026 E mail support xavant com Web www xavant com...