Specification
MAIN OUTPUTS
Voltage Range:
0V to 18V minimum (XPL 18-3)
0V to 30V minimum (XPL 30-1, XPL 30-2, XPL 30-2D, XPL 30-2T)
0 to 56V minimum (XPL 56-1)
Current Range:
0A to 1A minimum (XPL 30-1)
0A to 1.1A minimum (XPL 56-1)
0A to 2A minimum (XPL 30-2, XPL 30-2D, XPL 30-2T)
0A to 3.3A minimum (XPL 18-3)
Voltage Setting:
By coarse and fine controls.
Current Setting:
By single logarithmic control.
Operating Mode:
Constant voltage or constant current with automatic cross-over.
Output Switch:
Electronic. Preset voltage and current displayed when off.
Output Terminals:
Universal 4mm safety binding posts on 19mm (0.75”) pitch.
Output Impedance:
Typically <5m
Ω
in constant voltage mode.
Typically >50k
Ω
in constant current mode.
Output Protection:
Outputs will withstand forward voltages of up to 20V above the rated
output voltage. Reverse protection by diode clamp for currents up to 3A.
Load Regulation:
<0.01% of maximum output for 90% load change.
Line Regulation:
<0.01% of maximum output for 10% line change.
Ripple & Noise:
(20MHz bandwidth)
Typically <1mVrms (CV mode).
Transient Load Response: <20µs to within 50mV of set level for 90% load change.
Temperature Coefficient:
Typically <100ppm/°C.
Status Indication:
Output on lamp. Constant current mode lamp.
METER SPECIFICATIONS (Main Outputs)
Meter Types:
Dual 3 digit meters with 14mm (0.56") LEDs. Reading rate 4 Hz.
Meter Resolutions:
100mV, 10mA.
Meter Accuracies:
Voltage 0.3% of reading ± 1 digit.
Current 0.6% of reading ± 1 digit.
AUXILIARY LOGIC OUTPUT (XPL30-2T only)
Voltage:
Variable <1.5V to >5V by front panel control.
Meter Voltage Accuracy:
0.3% ± 1 digit.
Current Limit:
2A minimum.
Output Protection:
Output will withstand up to 7V forward voltage. Diode clamp reverse
protection for currents up to 3A.
Ripple & Noise
(20MHz bandwidth):
Typically <5mVrms, <15mVpk-pk (CV mode).
Load Regulation:
<3% for 90% load change.
Line Regulation:
<0.2% for 10% line voltage change.
Status Indication:
UNREG lamp.
3