Operation
Error Codes
Release 1.2
47
Error Codes
If the ERR flag in the accumulated status or fault registers has been activated, an
ERR? query will return an error number which corresponds to an event described
in the following table. The ERR? query will also clear the ERR bit in the register.
Table 3.11
Error Codes
ERROR # ERROR IDENTIFICATION
EXPLANATION
0
No Errors
4
Unrecognized Character
Received a character such as @,*,$.
Improper Number
Received a numeric character but the
characters were not a proper number.
Example: VSET,±10.3
Unrecognized String
Received an invalid command.
Syntax Error
Received an incorrectly placed word,
number, separator, or terminator.
Example: OFF SRQ, VOUT 6, MASK,
ERR
5
Number Out of Range
Specified a value for the command
which was outside of the allowed range.
6
Attempt to Exceed Soft Limits Attempted to program a voltage or
current greater than the soft limit.
Example: VMAX 500; VSET 550 LF
7
Improper Soft Limit
Attempted to program a soft limit less
than the output value.
8
Data Requested without a
Query Being Sent
The controller requested data from the
power supply without first sending a
query command.
9
OVP Set Below Output
Sent an OVSET command with a trip
value lower than the output voltage.
10
Slave Processor Not
Responding
The interface PCB slave processor did
not respond.
12
Illegal Calibration
Attempted calibration when the supply
was not in calibration mode. See
CMODE command.
Содержание GPIB-XPD
Страница 1: ...Operating Manual Internal GPIB Interface for XPD Series Programmable DC Power Supply GPIB XPD...
Страница 2: ......
Страница 3: ...Operating Manual for Internal GPIB Interface for XPD 500 Watt Series Programmable DC Power Supply...
Страница 10: ...About This Manual viii Operating Manual for GPIB for XPD Series Power Supply...
Страница 16: ...Features and Specifications Specifications 14 Operating Manual for GPIB for XPD Series Power Supply...
Страница 57: ......