M A - T 6 / M A - T 1 2 S P E C T R O P H O T O M E T E R
16
Language
This option is used to set the language that is displayed on the instrument during
operation.
To set the language, tap
Language
and then tap the desired language. A check mark appears
next to the selected language.
Calibration
This option is used to activate the calibration procedure.
To access, tap
Calibration.
Refer to the Calibration mode section later in this manual for detailed
information on performing a calibration.
Measurement Options
This option is used to set auto measurement mode and measurement averaging.
Auto Measure:
When set to On,
this option allows the instrument to automatically take a
measurement once all three pressure sensor indicator
s turn green in the display. No
measurement switch press or screen tap is required. Once a measurement is taken, the
instrument must be repositioned in order for another measurement to occur.
This icon appears on the targeting screen when Auto is set.
Measure Modes:
If available, this option is used to set the number of measurements required
for calculating a single measurement. Measurements are taken at different locations on a sample
to achieve average measurement values. The ava
ilable settings are No Average
(default),
Average 3, Average 5, and SMC n:m.
SMC n:m (Statistical Measurement Control) is a method of performing a statistical analysis of
several measurements to determine the quality of the measurements and/or the sample, before
an average value is calculated. A statistical analysis of the measurements' mean and standard
deviations eliminates outliers and determines the variability of the measurements. The number of
required measurements (3 to 15) is set along with the maximum number of measurements (3 to
15) that can be taken to achieve an average value.
To access, tap
Measurement Options
and then tap No Average (default), Average 3, Average 5,
or SMC n:m. When selecting SMC n:m, you will also need to set the number of good
measurements required and maximum measurements. Slide the dot icons to the right or left to
increase or decrease the measurement numbers.
Wi-Fi
This option is used to view all available networks when activated in Nucleos
Auto
QC.