CAPSURE Cosmetic User Manual
36
Cleaning the Calibration Reference
The calibration tile is imbedded in the underside of the white reference slider.
1.
Opening reference slider to a halfway position.
2.
Blow short bursts of clean, dry air onto the calibration tile. This should remove any
accumulated debris.
Calibration tile
Canned air straw