
page 25 of 84
ABS.ZERO
(BlueMETER
BASIC only)
Setting of absolute ZERO with a reversal measurement.
Using the reversal measurement is a simple way to determine the exact ZERO OFFSET of the
instrument as well as the exact inclination of the surface where the instrument is placed.
•
Slide the instrument onto a flat, horizontally levelled surface (e.g. engineer's surface
plate).
•
The position of the instrument is to be marked on the surface.
•
Use the key <ON/MODE> to get to the menu point [ABS.ZERO] and confirm with
<ENTER>
On the BlueMETER BASIC the following display is seen:
•
Use the zapper or confirm with <ENTER>. On the display the progress of the
collected values are visible in graphic form.
When the curve is getting a straight line confirm the first value with the zapper or with
<ENTER>
•
After the first reading the following display is seen:
This display means the first reading was successful. The instrument must be turned
180 degrees without lifting and slid exactly onto the previously marked position
•
Use the zapper or confirm with <ENTER>. On the display the progress of the
collected values are again visible in graphic form.
Содержание BlueLEVEL
Страница 2: ......
Страница 68: ...page 68 of 84 L FLOWCHARTS BLUEMETER BASIC ONLY L1 POWER UP...
Страница 69: ...page 69 of 84 L2 SAMPLING...
Страница 70: ...page 70 of 84 L3 SENSOR...
Страница 71: ...page 71 of 84 L4 REFRESH...
Страница 72: ...page 72 of 84 L5 RELATIVE ZERO REL ZERO...
Страница 73: ...page 73 of 84 L6 UNIT...
Страница 74: ...page 74 of 84 L7 ABSOLUTE ZERO ABS ZERO...
Страница 75: ...page 75 of 84 L8 FILTER...
Страница 76: ...page 76 of 84 L9 RADIO TRANSMISSION...
Страница 77: ...page 77 of 84 L10 JOIN...
Страница 78: ...page 78 of 84 L11 LEAVE...
Страница 79: ...page 79 of 84 L12 HOLD...