page 19 of 65 pages
During the measurement the display will graphically show the
current measurement.
Confirm the first measurement with the key
or with
the zapper. After 15 seconds the value will automatically be read.
After a successful reading of the first measuring value the position
of the instrument for the second measurement will appear in the
display.
Put the BlueLEVEL-2D now to the second position (turn the
instrument by 180 degrees in the horizontal) and press the key
again for reading the second measuring value.
Alternatively the second measuring value can also be entered with
the zapper.
During the measurement the display will graphically show the
current measurement.
Confirm the second measurement with the key
or
with the zapper. After 15 seconds the value will automatically be
read.
After termination of the reversal measurement the display for the
actual measurement under consideration of the ZERO OFFSET
will appear on the screen.
Notice:
The value of the "ZERO OFFSET" determined by a reversal measurement corresponds to the deviation
of the zero point of the BlueLEVEL-2D compared to the absolute Zero (gravity). The displayed
measuring value corresponds to
VALUE displayed =
Value of the BlueLEVEL-2D minus "ZERO OFFSET"
The reversal measurement described above should be repeated periodically in order to a chieve a high
measuring precision, particularly when the BlueLEVEL -2D has not been in use for a longer period.
Содержание BlueLEVEL-2D
Страница 2: ......
Страница 37: ...page 37 of 65 pages APPENDICES A FLOAT CHARTS GENERAL A1 START THE INSTRUMENT ON OFF...
Страница 38: ...page 38 of 65 pages A2 KEY REL ZERO...
Страница 39: ...page 39 of 65 pages A3 KEY HOLD FREEZING OF A MEASURING VALUE HOLD...
Страница 40: ...page 40 of 65 pages A4 MENU SELECTION...
Страница 41: ...page 41 of 65 pages A5 ADJUSTMENT OF THE DISPLAY DISPLAY...
Страница 42: ...page 42 of 65 pages A6 MEASUREMENT ABSOLUTE MODE ABSOLUTE...
Страница 43: ...page 43 of 65 pages A7 MEASURING ABSOLUTE CONSIDERING THE ZERO OFFSET MODE ABSOLUTE ZERO...
Страница 44: ...page 44 of 65 pages A8 MEASURING RELATIVE MODE REL ZERO...
Страница 45: ...page 45 of 65 pages A9 ADJUSTMENT OF THE MEASURING UNIT UNIT...
Страница 46: ...page 46 of 65 pages A10 SETTING OF LIMITS LIMITS...
Страница 47: ...page 47 of 65 pages A11 ADJUSTMENT OF FILTER TYPES FILTER...
Страница 48: ...page 48 of 65 pages A12 GROUPING OF INSTRUMENTS WITH WIRELESS TRANSMISSION JOIN...
Страница 49: ...page 49 of 65 pages A13 LEAVING A GROUP OF INSTRUMETNS LEAVE...
Страница 50: ...page 50 of 65 pages B FLOAT CHARTS OPTIONS B1 PIN CODE QUERY FOR ACCESS TO THE OPTIONS...
Страница 51: ...page 51 of 65 pages B2 SET PIN CODE...
Страница 52: ...page 52 of 65 pages B3 DISPLAY SETTINGS...
Страница 53: ...page 53 of 65 pages B4 LOGARITHMIC SCALE ON OFF LOGSCALE ON OFF...
Страница 54: ...page 54 of 65 pages B5 PROGRAMMABLE KEYS REL ZERO AND ZERO SELECT PROGRAMMABLE KEYS...
Страница 55: ...page 55 of 65 pages B6 SWITCHING FUNCTIONS ON OR OFF FUNCTIONS ON OFF...
Страница 56: ...page 56 of 65 pages B7 SWITCHING THE HIDING OF DISABLED FUNCTIONS ON OR OFF HIDE DISABLED FUNCTIONS ON OFF...
Страница 57: ...page 57 of 65 pages B8 SWITCH THE WIRELESS TRANSMISSION ON OR OFF RADIO ON OFF...
Страница 58: ...page 58 of 65 pages B9 SWITCH THE COMPENSATION OF GRAVITY ON OR OFF GRAVITATION COMPENSATION...
Страница 59: ...page 59 of 65 pages B10 DISPLAY THE FIRMWARE VERSION VERSION FIRMWARE...
Страница 60: ...page 60 of 65 pages B11 RESET TO FACTORY SETTINGS FACTORY RESET...
Страница 61: ...page 61 of 65 pages B12 FUNCTION CHECK DEVICE CHECK...