Ion concentration
pH/ION 7320
60
ba75935e04
12/2017
11.
Select
Continue
with
<
> <
>
and start the measurement with
<MENU/ENTER>
.
When the measurement is finished an entry window appears.
12.
Add the sample to the standard solution.
13.
Using
<
> <
>
and
<MENU/ENTER>
, enter the value for the volume
of the sample (
Sample volume
).
14.
Select
Continue
with
<
> <
>
and start the measurement with
<MENU/ENTER>
.
The measurement result is displayed when the measurement is com-
pleted.
15.
If necessary, start measuring further samples with
<MENU/ENTER>
.
Repeat steps 6 - 12 for all samples.
16.
Terminate the measuring method with
<M>
.
A safety query appears.
17.
Select
Yes
with
<
> <
>
.
18.
Confirm
Yes
with
<MENU/ENTER>
.
Measurement with the selected method is completed.
i
Add sample!
Sample volume
1.0 ml
Continue
Sample addition
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Sample addition
mg/l
USB output