pH/ION 7320
pH value
ba75935e04
12/2017
33
Continuing with
three- to five-point
calibration
14.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
15.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
16.
If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
pH
0.0
24.8
°C
AutoCal TEC
7.000
12.12.2012
12:12
AR
Buffer
mV
17.
Thoroughly rinse the electrode with deionized water.
18.
Immerse the electrode in the next buffer solution.
19.
When measuring without temperature sensor:
Enter the temperature of the buffer with
<
><
>
.
20.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
pH
180.0
24.8
°C
AutoCal TEC
4.010
12.12.2012
12:12
AR
Buffer
mV
Содержание pH/ION 7320
Страница 2: ...2 ba75935e04 12 2017 pH ION 7320 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pH ION 7320 6 ba75935e04 12 2017 ...
Страница 86: ...Transmitting data USB interface pH ION 7320 86 ba75935e04 12 2017 ...
Страница 92: ...What to do if pH ION 7320 92 ba75935e04 12 2017 ...
Страница 103: ......