ORP voltage
pH/ION 7320
42
ba75935e04
12/2017
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
Manually starting
the
stability control
You can start the
Stability control
manually at any time, irrespective of the set-
ting for automatic
Stability control
(see section 9.3.1 S
, page 75) in the
System
menu.
Freezing the
measured value
(HOLD function)
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
Measured parame-
ter
Time interval
Stability in the time interval
ORP voltage
15 seconds
∆
: better than 0.3 mV
Temperature
15 seconds
∆
: better than 0.5 °C
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2.
Using
<ENTER>
, activate the
Stability control
function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears and the display of the
measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<ENTER>
at any time. If the
Stability control
function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.
Using
<ENTER>
, start a further measurement with
Stability control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
Содержание pH/ION 7320
Страница 2: ...2 ba75935e04 12 2017 pH ION 7320 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pH ION 7320 6 ba75935e04 12 2017 ...
Страница 86: ...Transmitting data USB interface pH ION 7320 86 ba75935e04 12 2017 ...
Страница 92: ...What to do if pH ION 7320 92 ba75935e04 12 2017 ...
Страница 103: ......