pH value
pH/ION 7320
38
ba75935e04
12/2017
6.2.6 Calibration data
Displaying the
calibration data
The calibration record of the last calibration is to be found under the menu item,
<MENU/ENTER>
/
Calibration
/ /
Calibration record
. To open it, press the
<CAL__>
key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with
<F2>
[USB output]
.
Displaying the
calibration data
storage
The calibration records of the last calibrations (up to 10) are given in the menu,
<MENU/ENTER>
/
Calibration
/
Calibration data storage
.
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.
Содержание pH/ION 7320
Страница 2: ...2 ba75935e04 12 2017 pH ION 7320 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pH ION 7320 6 ba75935e04 12 2017 ...
Страница 86: ...Transmitting data USB interface pH ION 7320 86 ba75935e04 12 2017 ...
Страница 92: ...What to do if pH ION 7320 92 ba75935e04 12 2017 ...
Страница 103: ......