Operation
pH 315i
24
4.3.2
AutoCal DIN
For this procedure, use two or three different standard buffer solutions
according to DIN 19266 in ascending or descending order (type A, C,
D or F with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).
Note
The buffer solutions are automatically recognized by the measuring in-
strument. Depending on the instrument setting (see section
4.4 C
ONFIGURATION
), the instrument displays the relevant buffer nomi-
nal value or the current electrode voltage in mV.
1
Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol dis-
plays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).
2
Immerse the pH electrode in the first buffer solution.
3
Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
4
When the measured value is stable, Ct2 appears.
AutoCal TEC
CalError
ARng AR
LoBat
8
mV/pH
8
1
U
%
°F
S
ConCal
Auto Store
Time
Baud
Day.Month No.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
8
4
2
1
C d
°C
TP
pH
AutoCal DIN
U
AutoCal TEC
AutoCal TEC
mV
U
mV/pH
mol/L
8
Auto Store
TempError
RCL
LoBat
CalError
ARng
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
%
°F
S
mg/L
% ppm
ISE
8
4
2
°C
TP
mV/pH
mol/L
8
Auto Store
TempError
RCL
LoBat
CalError
ARng
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
%
°F
S
mg/L
% ppm
ISE
8
4
2
6
0
°C
TP
7 7
7
pH
mV
AR
0
AR
AutoCal DIN
AutoCal DIN
pH
1
4
Содержание pH 315i
Страница 4: ...List of contents pH 315i 4...
Страница 12: ...Commissioning pH 315i 12...
Страница 36: ...What to do if pH 315i 36...
Страница 44: ...Index pH 315i 44...