Operation
pH 315i
14
4.2
Measuring
Preparatory activities
Perform the following preparatory activities when you want to measure:
Note
Incorrect calibration of pH electrodes leads to incorrect measured val-
ues. Calibrate regularly before measuring.
Temperature sensor
Basically, pH measurements should only be performed using a temper-
ature probe. The temperature probe is shown on the display by TP.
If you nevertheless use an electrode without a temperature probe, the
instrument operates with a reference temperature of 25 °C. Thus, the
sample and calibration solution must have their temperatures adjusted
to 25 °C in order to avoid larger measurement errors.
Note
The pH meter automatically recognizes the type of temperature probe
used. This enables electrodes to be connected with the NTC30 or
Pt1000.
1
Connect the electrode to the measuring instrument.
2
Calibrate or check the measuring instrument with the elec-
trode.
3
Select the measuring mode with <M>.