Service menu
70
7.4.5
Maximum suction time
Setting the maximum suction time protects the membrane pump
against damage caused by permanent operation.
Example: If, because of a leaking suction hose, no water is taken in, the
suction procedure is automatically terminated after the maximum
suction time has expired.
Note
If the first sampling effort has taken in no water after the maximum
suction time is over, the suction procedure is automatically terminated.
If also the second sampling effort does not take in water, this causes
the ERROR CLOGGED error message (see section 9.1 E
).
The running program is not interrupted by the error.
1
Call up the service menu (see section 7.4.1 E
2
Using
ud
scroll to the MAX. SUCTION TIME menu.
3
Press
g
.
A cursor appears on the displayed value.
4
Using
ud
and
lr
increase or decrease the value.
5
Using
g
confirm the value.
The new value has been set; the cursor has disappeared.
6
Using
ud
switch to the next setting.
Содержание PB 150
Страница 1: ...ba74107e03 03 01 Sampler PB 150 Operating manual Sampler PB 150 ...
Страница 6: ...List of contents 6 ...
Страница 18: ...Overview 18 ...
Страница 22: ...Safety 22 ...
Страница 30: ...Principle operation 30 ...
Страница 50: ...Programming the automatic sampling 50 ...
Страница 90: ...Technical data 90 ...
Страница 93: ...Appendix 93 Changing the program continuation Main menu TIME RELATED SAMPLE RELATED YES NO VERZÖGERT AT ONCE ...
Страница 95: ...Appendix 95 11 1 6 Service menu Main menu ...
Страница 101: ...Appendix 101 11 4 Dimensions of stationary sampler PB 150 ...
Страница 106: ...Appendix 106 ...