Programming the automatic sampling
45
Fixed time for bottle
change
The fixed time for bottle change causes a bottle to be filled at a certain
time regardless of the time the program started.
When the time for the bottle change is fixed at 8:00 a. m. and the filling
time per bottle is 120 min, the bottles are filled at the following times:
In this example, the program starts at 09:45 a. m. and the bottle is
already changed for the first time after 15 min at 10:00 a. m.
Next bottle change
Here you specify at which time the first bottle will be filled (with FIXED
TIME FOR BOTTLE CHANGE only).
Bottle 1:
08:00 a. m. - 10:00 a. m.
Bottle 2:
10:00 a. m. - 12:00 p. m.
etc.
etc.
6
Press
g
.
The cursor appears in front of the selection.
7
Using
ud
select YES or NO.
8
Using
g
confirm YES or NO.
– NO: The BOTTLE VERSION menu opens (continue with
– YES: The NEXT BOTTLE CHANGE menu opens.
9
Press
g
.
A cursor appears on the displayed value.
10
Using
ud
and
lr
increase or decrease the value.
11
Using
g
confirm the displayed value. The BOTTLE VERSION
menu opens (continue with B
Содержание PB 150
Страница 1: ...ba74107e03 03 01 Sampler PB 150 Operating manual Sampler PB 150 ...
Страница 6: ...List of contents 6 ...
Страница 18: ...Overview 18 ...
Страница 22: ...Safety 22 ...
Страница 30: ...Principle operation 30 ...
Страница 50: ...Programming the automatic sampling 50 ...
Страница 90: ...Technical data 90 ...
Страница 93: ...Appendix 93 Changing the program continuation Main menu TIME RELATED SAMPLE RELATED YES NO VERZÖGERT AT ONCE ...
Страница 95: ...Appendix 95 11 1 6 Service menu Main menu ...
Страница 101: ...Appendix 101 11 4 Dimensions of stationary sampler PB 150 ...
Страница 106: ...Appendix 106 ...