Programming the automatic sampling
36
5.1.5
Sampling pause
During the specified SAMPLING PAUSE no sample is taken. With the
TIME RELATED BOTTLE FILLING, this means that the time between
2 bottle changes is increased.
The interval between two bottle fillings (INTERVAL) is only affected by
this if a filling is to take place in a Sampling pause. In this case, the next
filling takes place after the sampling pause.
With the SAMPLE RELATED BOTTLE FILLING, the SAMPLING
PAUSE has no function.
Calculation
Sampling pause = Interval between bottle changes - Filling time per bottle
In the SAMPLING PAUSE menu, enter 5 Min.
No sampling takes place in the first 5 minutes after each bottle change.
Given:
Number of sample bottles:
Bottle volume:
Time between bottle changes:
Filling time per bottle:
Interval:
24
1000 ml
65 min
60 min
8 min
Calculated:
Sampling pause:
65 min - 60 min =
5 min