Settings
Oxi 3310 IDS (Oxi 3315)
34
ba75944e02
10/2014
6.2
Sensor-independent settings
6.2.1
System
To open the
Storage & config
menu, press the
<ENTER_>
key in the measured
value display. After completing the settings, switch to the measured value dis-
play with
<M>
.
Default settings are printed in
bold
.
Salinity
0.0
... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item if only available if the
manual salinity correction (
Sal cor-
rection
) isswitched on.
Response time t90
30 ... 300 s
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces
the limits of variation of the mea-
sured value. The signal filter is char-
acterized by the response time t90.
This is the time after which 90 % of a
signal change is displayed.
Reset
-
Resets all sensor settings to the
delivery condition (see section 6.3.1
R
Menu item
Possible
setting
Explanation
Menu item
Possible setting
Explanation
System
/
General
/
Lan-
guage
Deutsch
English
(more)
Selects the menu language
System
/
General
/
Audio
signal
on
off
Switches on/off the beep on keystroke
System
/
General
/
Illumi-
nation
Auto
on
off
Switches the display illumination on/off
System
/
General
/
Con-
trast
0 ...
50
... 100 %
Changes the display contrast
System
/
General
/
Shutoff
time
10 min ...
1h
...
24 h
Adjusts the switch-off time
System
/
General
/
Tem-
perature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed with the
selected unit.