Dissolved oxygen
Oxi 3310 IDS (Oxi 3315)
30
ba75944e02
10/2014
5.3.5 Calibration with
Comparison meas.
(FDO Comp)
For this calibration procedure, the
Comparison meas.
setting must be set to
on
in the
Calibration
menu.
Proceed as follows to calibrate the instrument:
Before calibrating via a comparison measurement, the sensor
should be calibrated in the air calibration vessel.
1.
Connect the D.O. sensor to the meter.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with
<CAL>
.
4.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement or accept the calibration
value with
<ENTER>
.
The multiplier that was set last is displayed.
O
2
7.92
mg/l
25.1
°C
FDO Comp
22.09.2008
08:00
O
2
7.92
mg/l
0.973
FDO Comp
22.09.2008
08:00