Multi 9310
Settings
ba75937e03
12/2011
67
Calibration
/
Cali-
bration interval
1 ...
180
...
999 d
Calibration interval
for the D.O. sensor
(in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Calibration
/
Com-
parison meas.
on
off
Enables to adjust the measured value
with the aid of a comparison measure-
ment, e.g. Winkler titration.
For details, see section 7.3 C
FDO Check
/
Start FDO Check
-
Starts the FDO
®
check procedure
FDO Check
/
Check interval
1 ...
60
...
999 d
Interval for the
FDO Check
(in days).
The meter reminds you to check the
sensor regularly by
FDO Check
status
indicator in the measuring screen.
Sal correction
on
off
Manual salt content correction for con-
centration measurements.
Salinity
0.0
... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
manual salt content correction is
switched on.
Response time
t90
30
... 300 s
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces the
limits of variation of the measured
value. The signal filter is characterized
by the response time t90. This is the
time after which 90 % of a signal
change is displayed.
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 9.6.1
R
Menu item
Possible
setting
Description