Multi 9310
Conductivity
ba75937e03
12/2011
61
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(USB output)
Menu item
Setting/
function
Description
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration records.
Further options:
z
Scroll through the calibration
records with
<
S
S
S
S
><
T
T
T
T
>
.
z
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
z
Output all calibration records to the
interface with
<F2__>
[USB out-
put]
.
z
Quit the display with
<F1>
/[Back]
or
<ENTER>
.
z
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data storage
to the interface
Display
Calibration record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 14
W
CALIBRATION Cond
Calibration date 01.07.2011 07:43:33
TetraCon 925
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor
+++