7.2.1 Why should you check the sensor?
With the FDO
®
Check procedure, you can find out in a simple manner
whether the FDO
®
925 D.O. sensor should be cleaned or calibrated.
7.2.2 When should you check the sensor?
Checking can be useful in the following cases:
z
When the check interval has expired
(The [check] status indicator is displayed.)
z
If the measured values seem to be implausible
z
If you assume that the sensor cap is contaminated or at the end of its lifetime
z
After the sensor cap was exchanged
z
Routinely within the framework of the company quality assurance
7.2.3 Carrying out the FDO® check procedure
FDO
®
check
procedure
Check in water vapor-saturated air.
Use the check and storage beaker (FDO
®
Check) to carry out the FDO
®
check
procedure.
Stability control
(AutoRead)
In the FDO
®
check procedure, the Stability control function (AutoRead) is auto-
matically activated.
Proceed as follows to carry out the FDO
®
check procedure:
1.
Connect the D.O. sensor to the meter.
2.
Place the D.O. sensor in the check and storage beaker, FDO
®
Check.
The sponge in the check and storage beaker must be moist (not
wet). Leave the sensor in the check and storage beaker long
enough so it can adapt to the ambient temperature.
3.
Start the FDO
®
check procedure in the measuring menu with
FDO
Check / Start FDO Check
.
The meter switches to the measured parameter, %.