KANE3500 

PRESSURE METER 

 
 
 
 
 
 

 

 

 
Stock No : 18853 

          January 2010

 

 

 

 

 

KANE3500 

PRESSURE METER 

 
 
 
 
 
 

 

 
Stock No : 18853 

          January 2010

 

 

 

Содержание KANE3500

Страница 1: ...KANE3500 PRESSURE METER Stock No 18853 January 2010 KANE3500 PRESSURE METER Stock No 18853 January 2010 ...

Страница 2: ...a can be live data frozen data or stored data 255 sets of tests can be stored in non volatile memory GENERAL OPERATION When the instrument is first switched on using the top line of the display shows in sequence the date the time the model version number the battery status and the pressure units The bottom line of the display shows a countdown number that starts at 10 and reduces to zero At zero t...

Страница 3: ... The meter counts down from 10 after OFF is pressed Press to cancel OFF MENU Press and hold for MENU Two lines of 20 characters can be added to the header of printouts The meters are controlled using 8 buttons The eight buttons are ON OFF Press for ON or OFF The meter counts down from 10 after OFF is pressed Press to cancel OFF MENU Press and hold for MENU ...

Страница 4: ...CKLIGHT HOLD PRINT A quick press to HOLD the reading and the display flashes A longer press to PRINT ZERO Press and hold until there is a beep to ZERO the reading BACKLIGHT Press and hold until there is a beep to toggle the BACKLIGHT HOLD PRINT A quick press to HOLD the reading and the display flashes A longer press to PRINT ...

Страница 5: ... arrows also change settings such as date time and other menu items when MENU mode has been selected UP DOWN Press either to change the bottom line of the display ENTER Press to cancel OFF The buttons with and arrows also change settings such as date time and other menu items when MENU mode has been selected ...

Страница 6: ...ery Set the meter s correct time date and units These are then stored when the meter is switched off 1 BEFORE USING THE METER FOR THE FIRST TIME Remove the cover and fit a new battery in the battery compartment Take great care to ensure that the battery is fitted with the correct battery polarity Then replace the battery cover Always check that the meter is working correctly after replacing the ba...

Страница 7: ... measure Excessive pressure 5 times nominal range can permanently damage the meter s pressure sensor 2 BEFORE USING THE METER EVERY TIME After switch on check that date and time are correct and battery power is sufficient and the displayed units are the ones required SAFETY WARNING Never connect to a pressure source where you are not sure what the maximum pressure might be Always ensure that the m...

Страница 8: ...e input measurements of pressure or vacuum Only use the input when taking a differential measurement NOTE Use the input for all single input measurements of pressure or vacuum Only use the input when taking a differential measurement ...

Страница 9: ...ess and use the button to select the following function for change STORE SETUP PRESSURE SERVICE EXIT 3 THE MENU FUNCTIONS The MENU structure comprises main menu options STORE SETUP and PRESSURE have a sub menu structure SERVICE is for use by an approved service organisation only MENU Press and use the button to select the following function for change STORE SETUP PRESSURE SERVICE EXIT ...

Страница 10: ...he key and change using the keys the function The final logical returns you to the main menu display Press to select a main menu function When you have selected the function to change press to select Repeat this to scroll through the menu and select using the key and change using the keys the function The final logical returns you to the main menu display ...

Страница 11: ...unction the final logical is pressed No changes are made if the MENU function is aborted by pressing To exit the MENU function the final logical is pressed No changes are made if the MENU function is aborted by pressing ...

Страница 12: ...of decimal places LOW one less decimal place displayed PS UNITS mBAR inH2O hPa mmHg PSI kPa Pa mmH2O PRESSURE SMOOTH OFF Normal ON Readings filtered to damp out fluctuations RESOLVE HIGH maximum number of decimal places LOW one less decimal place displayed PS UNITS mBAR inH2O hPa mmHg PSI kPa Pa mmH2O ...

Страница 13: ...ime 2 Stabilisation Period Time 3 Tightness Test duration Set using Times are in minutes 01 1 Minute REPORT TEST VIEW DEL ALL SETTING NB Automated Let By Test Stabilisation Tightness Test Display stored report results max 8 Delete all test results Time 1 Let By Test duration Time 2 Stabilisation Period Time 3 Tightness Test duration Set using Times are in minutes 01 1 Minute ...

Страница 14: ...e user defined format Header 2 lines each of up to 20 characters that appear on the printout Use and SETUP 1 2 3 4 5 Temperature units o F or o C AUTO OFF adjusts the auto power off time A setting of 00 sets the meter for manual switch off only Time Uses Military time as standard 7am 07 00 7pm 19 00 Date user defined format Header 2 lines each of up to 20 characters that appear on the printout Use...

Страница 15: ...6 7 8 Language English French Italian German Spanish The display s contrast EXIT 6 7 8 Language English French Italian German Spanish The display s contrast EXIT ...

Страница 16: ...to exit this mode LOG Saves a set of results when is pressed Storage capacity is 255 sets of results STORE VIEW The test number appears on the top line Use or to change test number The bottom line shows the measured value Use to exit this mode LOG Saves a set of results when is pressed Storage capacity is 255 sets of results ...

Страница 17: ... seconds and then press to activate and exit DEL ALL Clears all the memory when YES and pressed EXIT AUTO STO Automatically stores readings at a preset interval Press then select YES or NO If YES set the desired time interval between 2 and 99 seconds and then press to activate and exit DEL ALL Clears all the memory when YES and pressed EXIT ...

Страница 18: ...E OPTIONS Max Pressure Min Pressure Time Date Pressure units Battery Status Instrument Internal Temp BOTTOM LINE OPTIONS Max Pressure Min Pressure Time Date Pressure units Battery Status Instrument Internal Temp ...

Страница 19: ...Max Pressure is displayed as P 1 3 1 9 BOTTOM LINE OPTIONS Min Pressure is displayed as P 1 3 _ 1 1 Max Pressure is displayed as P 1 3 1 9 BOTTOM LINE OPTIONS Min Pressure is displayed as P 1 3 _ 1 1 ...

Страница 20: ... zero the meter in the position you are taking the readings and hold the meter still during the test 4 MEASURING Make sure you do not exceed the meter s operating specifications Do not exceed the meter s internal temperature operating range Do not put the meter on a hot surface When taking critical draft measurements always re zero the meter in the position you are taking the readings and hold the...

Страница 21: ...le channel measurements pressure or vacuum Only use the left hand port for differential measurements Differential Port Port 1 IR Printer Output Always use the top right hand port Port 1 for taking single channel measurements pressure or vacuum Only use the left hand port for differential measurements ...

Страница 22: ...s switched on for at least five minutes and then re zeroed before taking such sensitive measurements If the pressure being measured exceeds the meter s design range the display will show OR for over range When taking draft readings at very low pressure or draft levels for maximum accuracy re zero the meter in the orientation that it is being used This eliminated gravity effects on the very sensiti...

Страница 23: ... PRINTOUTS The standard printout is The MAX and MIN readings are those observed since last switch on EXAMPLE PRINTOUTS The standard printout is The MAX and MIN readings are those observed since last switch on ...

Страница 24: ...NOTE Printouts of stored readings will also include the TEST NO below the header NOTE Printouts of stored readings will also include the TEST NO below the header ...

Страница 25: ...in minutes TIME 01 Let By Test duration TIME 02 Stabilisation period TIME 03 Tightness Test duration 5 LET BY and TIGHTNESS TESTING only applicable for KANE3500 1 UK and KANE3500 2 UK To set the duration of each test press select PRESSURE select REPORT select SETTINGS Use and to set the times in minutes TIME 01 Let By Test duration TIME 02 Stabilisation period TIME 03 Tightness Test duration ...

Страница 26: ...nd adjust the pressure accordingly Once the correct pressure has been achieved press TIME 59 P2 10 05 LET BY TEST only applicable for KANE3500 1 UK and KANE3500 2 UK Press select PRESSURE select REPORT select TEST If the let by test is required change to YES and press CONNECT P1 0 00 Connect the meter to the system under test and adjust the pressure accordingly Once the correct pressue has been ac...

Страница 27: ...xit the test sequence If the system passes the let by test press to continue with the tightness test The display shows the countdown timer Time 1 in REPORT SETTINGS and the measured pressure Once the countdown has finished the display shows the initial pressure P1 and the final pressure P2 P1 10 05 P2 10 05 If the system fails the let by test press to exit the test sequence If the system passes th...

Страница 28: ...d adjust the pressure accordingly Once the correct pressure has been achieved press TIME 59 P3 19 95 TIGHTNESS TEST only applicable for KANE3500 1 UK and KANE3500 2 UK After completing the let by test or after selecting NO the stabilisation and tightness test can be performed The display shows CONNECT P 0 00 Connect the meter to the system under test and adjust the pressure accordingly Once the co...

Страница 29: ...ished P3 is stored and the display changes to TIME 120 P4 19 85 Where TIME Time 3 in REPORT SETTINGS The display shows the countdown timer Time 2 in REPORT SETTINGS and the measured pressure Once the countdown has finished P3 is stored and the display changes to TIME 120 P4 19 85 Where TIME Time 3 in REPORT SETTINGS ...

Страница 30: ...essure at the end of the test period Press to log the result The display shows the log number LOG 03 At the end of the countdown period the display automatically changes to P3 19 87 P4 19 84 Where P3 is the pressure at the start of the test period and P4 is the final pressure at the end of the test period Press to log the result The display shows the log number LOG 03 ...

Страница 31: ...to print if required The logged data can always be accessed by VIEW in the report menu Press to EXIT Press PRINT to print if required The logged data can always be accessed by VIEW in the report menu Press to EXIT ...

Страница 32: ...htness Test Printout NOTE Printouts of stored readings will also include the LOG NO Below the header Let By Test and Tightness Test Printout NOTE Printouts of stored readings will also include the LOG NO Below the header ...

Страница 33: ...tion is performed in firmware and there are no user accessible adjustments or user serviceable parts 6 METER ANNUAL RECALIBRATION AND SERVICE The meter should be re calibrated and serviced annually by a Kane approved service centre Local regulations may require more frequent re calibration Calibration is performed in firmware and there are no user accessible adjustments or user serviceable parts ...

Страница 34: ......

Страница 35: ... Pressure Ranges mBar mmH2O Pa kPa 80 00 800 0 8000 8 000 PSI mmHg hPa inH20 1 000 60 00 80 00 30 00 7 SPECIFICATION Model KANE3500 1 Nominal Pressure Ranges mBar mmH2O Pa kPa 80 00 800 0 8000 8 000 PSI mmHg hPa inH20 1 000 60 00 80 00 30 00 ...

Страница 36: ...ure Ranges mBar mmH2O Pa kPa 160 00 1600 0 16000 16 000 PSI mmHg hPa inH20 2 000 120 00 160 00 60 00 SPECIFICATION Model KANE3500 2 Nominal Pressure Ranges mBar mmH2O Pa kPa 160 00 1600 0 16000 16 000 PSI mmHg hPa inH20 2 000 120 00 160 00 60 00 ...

Страница 37: ... Pressure Ranges mBar mmH2O Pa kPa 400 0 4000 32000 40 00 PSI mmHg hPa inH20 5 000 300 0 400 0 150 0 SPECIFICATION Model KANE3500 5 Nominal Pressure Ranges mBar mmH2O Pa kPa 400 0 4000 32000 40 00 PSI mmHg hPa inH20 5 000 300 0 400 0 150 0 ...

Страница 38: ...NE3500 15 Nominal Pressure Ranges mBar kPa PSI 1040 0 100 00 15 00 mmHg hPa inH20 780 0 1040 0 400 0 SPECIFICATION Model KANE3500 15 Nominal Pressure Ranges mBar kPa PSI 1040 0 100 00 15 00 mmHg hPa inH20 780 0 1040 0 400 0 ...

Страница 39: ...E3500 30 Nominal Pressure Ranges mBar kPa PSI 2070 0 207 00 30 00 mmHg hPa inH20 1560 0 2070 0 800 0 SPECIFICATION Model KANE3500 30 Nominal Pressure Ranges mBar kPa PSI 2070 0 207 00 30 00 mmHg hPa inH20 1560 0 2070 0 800 0 ...

Страница 40: ...curacy 20 Pa 0 1 Pa 0 5 Pa 100 Pa 0 1 Pa 3 Pa 2000 Pa 1 Pa 3 of reading 80 hPa 0 01 hPa 3 of reading Pressure Measurement Model KANE3500 1 Max overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy 20 Pa 0 1 Pa 0 5 Pa 100 Pa 0 1 Pa 3 Pa 2000 Pa 1 Pa 3 of reading 80 hPa 0 01 hPa 3 of reading ...

Страница 41: ... range Range Resolution Accuracy 20 Pa 0 3 Pa 2 Pa 2000 Pa 0 3 Pa 5 Pa 160 hPa 0 01 hPa 3 of reading Pressure Measurement Model KANE3500 2 Max overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy 20 Pa 0 3 Pa 2 Pa 2000 Pa 0 3 Pa 5 Pa 160 hPa 0 01 hPa 3 of reading ...

Страница 42: ... Range Resolution Accuracy 32000 Pa 1 Pa 5 Pa 300 hPa 0 01 hPa 0 05 hPa 400 hPa 0 1 hPa 3 of reading Pressure Measurement Model KANE3500 5 Max overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy 32000 Pa 1 Pa 5 Pa 300 hPa 0 01 hPa 0 05 hPa 400 hPa 0 1 hPa 3 of reading ...

Страница 43: ...s nominal range Range Resolution Accuracy 309 99 hPA 0 01 hPa 10 hPa 1040 0 hPA 0 1 hPa 3 of reading Pressure Measurement Model KANE3500 15 Max overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy 309 99 hPA 0 01 hPa 10 hPa 1040 0 hPA 0 1 hPa 3 of reading ...

Страница 44: ... times nominal range Range Resolution Accuracy 1076 hPA 0 1 hPa 10 hPa 2070 0 hPA 1 hPa 3 of reading Pressure Measurement Model KANE3500 30 Max overrange without damage to sensor is 5 times nominal range Range Resolution Accuracy 1076 hPA 0 1 hPa 10 hPa 2070 0 hPA 1 hPa 3 of reading ...

Страница 45: ... x 80mm x 40mm 180mm incl spigots 0 C to 45 C 10 to 90 RH non condensing 9 Volt PP3 alkaline battery Dimensions Weight Handset Ambient Operating Range Power Supply 295 grams with battery 160mm x 80mm x 40mm 180mm incl spigots 0 C to 45 C 10 to 90 RH non condensing 9 Volt PP3 alkaline battery ...

Страница 46: ...tromagnetic radiation levels and occasionally it may be appropriate to check the meter before use by 8 ELECTROMAGNETIC COMPATIBILITY European Council Directive 89 336 EEC requires electronic equipment not to generate electromagnetic disturbances exceeding defined levels and have adequate immunity levels for normal operation Specific standards applicable to this meter are stated below As there are ...

Страница 47: ...position to minimize interference or switch off if possible the offending equipment during your test Use the normal start up sequence in the location where the meter will be used Switch on all localized electrical equipment capable of causing interference Check all readings are as expected A level of disturbance is acceptable If not acceptable adjust the meter s position to minimize interference o...

Страница 48: ...interference has occurred and this advice is only given to satisfy the requirements of the Directive At the time of writing this manual Jan 2010 Kane International Ltd are not aware of any field based situation where such interference has occurred and this advice is only given to satisfy the requirements of the Directive ...

Страница 49: ...ication EC EMC KI KANE3500 details the specific test configuration performance and conditions of use ELECTROMAGNETIC COMPATIBILITY This product has been tested for compliance with the following generic standards EN 61000 6 3 2001 EN 61000 6 1 2001 and is certified to be compliant Specification EC EMC KI KANE3500 details the specific test configuration performance and conditions of use ...

Страница 50: ...sent to the appropriate recycling centre in accordance with current legislation and local guidelines Please Note Batteries used in this instrument should be disposed of in accordance with current legislation and local guidelines At the end of its life the meter should be sent to the appropriate recycling centre in accordance with current legislation and local guidelines ...

Страница 51: ...and year The order can be changed using the menu function Date is recorded when each test is printed APPENDIX 1 Main Parameters UNITS DATE User selectable from PSI hPa inH2O mBar mmH2O Pa kPa NOTE Not all units available on all models Date shown as day month and year The order can be changed using the menu function Date is recorded when each test is printed ...

Страница 52: ... date and time for up to one minute if outside this time it may be necessary to re enter the details TIME The time is shown in hours and minutes expressed in Military time or the 24 hr clock Time is recorded when each test is stored printed NOTE When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be necessary to re ...

Страница 53: ...s than 10 of charge and should be replaced readings may be affected if used with low power batteries SYMBOLS used on the display T P The internal temperature of the meter The pressure reading in the user s selected units Displays the Battery power available in When the LO BAT symbol appears this indicates the batteries are less than 10 of charge and should be replaced readings may be affected if u...

Страница 54: ......

Страница 55: ......

Страница 56: ...ng the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specifications are subject to change without notice The user is responsible for determining the suitability of this product ...