Multi 3430
Operation
39
ba75865e02
12/2009
4.6.3
Settings for pH measurements
Overview
The following settings are possible for pH measurements:
z
Resolution
z
Calibration interval
z
Buffers for calibration
z
Automatic stability control
z
Unit for slope
z
Calibration record
(display)
Settings
The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required measured parameter in the measured value display and press
the
<MENU/ENTER>
key. After completing the settings, switch to the
measured value display with
<M>
.
Default settings are printed in
bold
.
Menu item
Possible set-
ting
Description
Calibration
/
Calibra-
tion record
-
Displays the calibration
record of the last calibration.
Calibration
/
Buffer
TEC
NIST/DIN
ConCal
...
Buffer sets to be used for pH
calibration.
For further buffers and
details, see section 4.6.4.
Calibration
/
Single-
point calibration
Yes
No
Quick calibration with 1 buf-
fer
Calibration
/
Calibra-
tion interval
1 ...
7
... 999 d
Calibration interval
for the
IDS pH sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Calibration
/
Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the
Nernst slope of
-59.2 mV/pH (100 x deter-
mined slope/Nernst slope).
Calibration
/
Calibra-
tion data storage
Displays the last calibration
records (max. 10)
Man. temperature
-25 ...
+25
...
+130 °C
Entry of the manually deter-
mined temperature.
Содержание Multi 3430
Страница 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Страница 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Страница 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Страница 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Страница 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Страница 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Страница 101: ......