4.6.2
Measuring the pH value
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.
You can start the
Stability control
manually at any time, irrespective of
the setting for automatic
Stability control
(see page 34) in the
Measure-
ment
menu.
1
Perform the preparatory activities according to section 4.6.1.
2
Immerse the IDS pH sensor in the test sample.
3
Select the pH or mV display with
<M>
.
30.09.2009 08:00
Info
1
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2
Using
<MENU/ENTER>
, activate the
Stability control
function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.
Содержание Multi 3430
Страница 6: ...Contents Multi 3430 6 ba75865e01 12 2009...
Страница 14: ...Overview Multi 3430 14 ba75865e02 12 2009...
Страница 18: ...Safety Multi 3430 18 ba75865e02 12 2009...
Страница 22: ...Commissioning Multi 3430 22 ba75865e02 12 2009...
Страница 86: ...Operation Multi 3430 86 ba75865e02 12 2009...
Страница 100: ...Appendix Firmware update Multi 3430 100 ba75865e02 12 2009...
Страница 101: ......