MIQ/CR3; DIQ/CR3
Indexes
21
ba64107d13
03/2018
7
Indexes
7.1
Explanation of the messages
In this chapter you will find a list with all the message codes and corresponding
message texts that may occur in the log book of the IQ S
ENSOR
N
ET
system for
the MIQ/CR3 output module.
7.1.1 Error messages
7.1.2 Info messages
The MIQ/CR3 or DIQ/CR3 output module does not send any info messages.
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
All message codes of the MIQ/CR3; DIQ/CR3 output module end
with the number "411".
Message code
Message text
EA4411
Current output range undercut
* Check process
* Check settings and, if necessary, change them
EA5411
Current output range exceeded
* Check process
* Check settings and, if necessary, change them
EI3411
Burden resistor too large (> 500 Ohm) or current loop interrupted
* Check burden, terminal connections and connection lines