Technical Data
pH 170/
pH 296
171
Technical data for pH 170
Display
Measured value
LCD 7 segment
•
pH / Redox
3-digit
•
Temperature
4-digit
Units, settings
Symbols
pH measurement
Signal input
Low impedance or high impedance
Display and measuring
range
0.00 to 14.00 pH
Measured value resolution
0.01 pH
Accuracy
± 0.01 pH ± 1 digit
Automatic temperature
compensation
•
NTC
–
5 to + 100
°
C
•
PT100/PT1000
–
20 to + 130
°
C
Reference temperature
+ 25
°
C
Calibration
•
AutoCal TEC
Automatic calibration with
technical buffers
•
AutoCal TEC
Automatic calibration with
rEC1
technical buffers and
subsequent output of the
buffer used in the calibration
•
ConCal
Manual calibration with
arbitrary buffer solutions
Calibrating range
•
Slope range
–
62 mV/pH to
–
50 mV/pH
•
Asymmetry potential
–
45 mV to + 45 mV