Operation
Cond 3210
26
ba75794e07
07/2017
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.4.3
Temperature compensation
The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20
or
Tr25
.
You can select one of the following temperature compensation meth-
ods:
2
Using
<ENTER>
, activate the
Stability control
function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.
Note
You can prematurely terminate the
Stability control
function manually
with
<ENTER>
at any time. If the
Stability control
function is prema-
turely terminated, the current measurement data are output to the
interface without the AutoRead info.
3
Using
<ENTER>
, start a further measurement with stability
control.
or
Release the frozen measured value again with
<AR>
or
<M>
.
The [AR] status display disappears. The display switches back
to the previous indication.
Measured
parameter
Time interval
Stability in the time interval
Conductivity
ϰ
10 seconds
Δ
: better than 1.0% of mea-
sured value
Temperature
15 seconds
Δ
: better than 0.5 °C