SI-MT/SI-MKB/SI-HTB
WORLD PRECISION IN STRU MENTS
41
TROUBLESHOOTING
Issue
Possible Cause
Solution
Chassis has no
power
One of the two power switches
is off.
Verify that the power switch one the
back of the chassis and the power switch
on the front panel are both in the on (I)
position.
The power cord is loose or not
connected properly to the AC
wall outlet
Unplug the power cord from the wall
and the chassis and re-install it.
S
I-BA
M21-L
C
B
has
no
output
signal
(0.0V D
C
)
Poor force transducer connec-
tion
Verify that the cables are securely con-
nected to the
SI-BAM21-LCB
.
BNC cable is bad
Try substituting a different BNC cable to
troubleshoot the cause.
Transducer failed
Try substituting a different force trans-
ducer to troubleshoot the cause.
Resonance
noise
still exists on the
transducer output
signal
Anti-oscillation frequency is not
set properly
Repeat the adjustment of the anti-
oscillation fi lter. See "Adjusting the Anti-
Oscillation Filter" on page 29. Verify
that the pulser amplitude is reduced
below maximum before trying another
anti-oscillation frequency.
NOTE
: If you have a problem/issue with that falls outside the defi nitions of this
troubleshooting section, contact the WPI Technical Support team at 941.371.1003 or
[email protected].