TROUBLESHOOTING
The PV850 continuously monitors the tank pressure to detect malfunctions in the
r ssur s st
n n on t au t t t
ff r nt rror
ssa s ar
displayed. When errors are detected an error screen with the error number and
suggested action is displayed. Upon exiting the Error Screen, the system returns to
the main screen with the Injection Port disabled. After addressing the failure source,
press the yellow banner to re-enable the port.
On power up the
PV850
performs a diagnostic self-test. If there is a failure during the
self-test, an error code displays and the unit stops the sequence. It will not progress
nto t o rat ona o un ss t s o r off an t n on an t s
t st
completes without error.
ERROR Condition
Possible Cause
Action
104
Tank pressure is too
low to deliver the
desired injection
pressure.
Tank depleted due to
previous injections or
leak.
Wait for pump to charge. Check
all pressure connections and
make sure the pipette is not
broken.
202
Failed to charge to the
maximum pressure
during the self-test.
Charging time
exceeded.
Ensure nothing is connected to
Injection Port. Call WPI technical
support if the problem persists.
203
High setting port valve
test failed during the
self-test.
Failed to reach the
proper Injection Port
pressure.
Ensure nothing is connected to
Injection Port. Call WPI technical
support if problem persists.
204
Low setting port valve
test failed during the
self-test.
Failed to reach the
proper Injection Port
pressure.
Ensure nothing is connected to
Injection Port. Call WPI technical
support if problem persists.
NOTE
ou a a ro
/ ssu t t at a s outs
t
n t ons o t s
trou
s oot n s t on onta t t P
n a Su ort t a at (
)
or
t
n a su ort
n o