10
World Precision Instruments
Gasket Color
Green Black
Red
White
P
tt a
t r (
)
1.0
1.2
1.5
1.65
!
WARNING: DANGER OF INJURY EXISTS IF THE PIPETTE IS INSECURE. HIGH
PRESSURE CAN CAUSE EJECTION AT HIGH VELOCITY.
5. The other end of the small tubing has a quick connector. Connect it with the blue
PV850
Tubing Assembly provided.
6. Verify that the Injection Port is disabled. If not, touch the Settings area of the display
to access the Settings menu and touch the Port ON button to set it to Port OFF.
7. Align the easy-connect Injector connector with the injector port on the front of
the
PV850
, slide the connector securely into the port and attach it with a slight
clockwise twist.
Fig. 12—(Left) Align the connector with the Injection Port, slide the connector in position and
rotate the connector a little clockwise to lock it into place.
Fig. 13—(Right) The PicoNozzle is connected to the injection port.
Connecting to a Pulse Generator for TTL Control
If you wish to use a computer or pulse generator to send a TTL pulse to trigger an
injection, connect the
803130
Y stereo splitter to the foot switch port on the rear
of the
PV850
. Connect the CBL102 jack to one of the female ports on the Y stereo
splitter. Plug the BNC cable into the appropriate TTL port.
803130
Y Stereo Splitter
CBL102
Stereo jack to BNC
To computer or
pulse generator
Connect to
Foot Switch
Port
Fig. 14—
Use the CBL102 and the 803130 to configure the system for a TTL trigger.