Chapter 5 Panel and Function
5.1 Function Panel
Display
Description
1
.
Press.
Real time in-chamber pressure
2
.
Temp.
Real time in-chamber temperature and error code
3
.
Start lamp
Flash when the program starts
4
.
Ster
ilize lamp
Flash in the process of sterilization
5
.
End lamp
Lighted up when the program ends
6
.
121℃ lamp
121℃ sterilizing temperature indicator
7
.
134℃ lamp
134℃ sterilizing temperature indicator
8
.
Open lamp
Lighted up when the release valve is open
manually
9
.
Close lamp
Lighted up when the release valve is closed
manually
10
.
Dry lamp
Flash when you choose drying process, Lighted up
when it ends
Controlling Button
Description
℃
Choose sterilizing temp. 121℃/134℃
Release
Open or close the release valve
DRY
Choose drying program
START
Start sterilizer program
5.2 Operation Process
1
、
Connect one end of the draining tube to the equipment and put the other end into the waste
water tank. Attention: In the drying process, in the chamber there may be a thermal vacuum, to
avoid the back-flow of the water in the waste water tank which will affect the drying effect, please
assure that the drain outlet is higher than the used water’s level in the waste water tank, a>5cm.
Used Water Level
a
2
、
Turn the two fore pads counterclockwise and evenly elevate the equipment to tilt the device
backwards to add more distilled water.