Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
47
of
87
No
Steps of procedure
Display and notes
33
Use the + and - keys to select the required
level of PASSWORD PROTECTION:
Settings and Access
Settings only
Off
Press
key when finished
If the SETTINGS AND ACCESS or the
SETTINGS ONLY option is selected the
following message will appear:
NOTE
Operator 01 is the Supervisor.
6666 is the default code, to which access codes are reset.
34
Set each code digit using the + and - keys
and the
and
keys to move the cursor
Press
key to finish when the cursor is on
the last digit
The following message will appear:
35
Set the code for Operators 02 to 08 in the
same way as for OPERATOR 01
NOTE
Unused operator codes should be
set to the same value as the Supervisor
code.
Press
key when all Operator codes have
been set
The following message will appear:
All settings are saved in non-volatile memory.
After a few seconds the Temperature and Level
display will reappear
If the password codes are lost or corrupted they may be reset to the default code 6666
using a PC and the PC Cable (M519CE) or the USB-PC Cable (M549CE).
PC software is available by E-mail for this purpose.
Instructions are supplied with the PC software.