Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
37
of
87
7.3 CONTROLLER CONFIGURATION AND CALIBRATION
Table 7
Configuration/Calibration Menu's options of the controller
Menu CONFIG/CALIBRATE
Language
Thermocouple calibration
Printer/PC connection and mode of operation.
Level Display Mode (High/Low/Normal or cm)
Temperature Control (On/off + timer settings)
Lid Switch Type (NO/NC)
Gas Bypass Setup (on/off)
Valve On/Off (No LN2)
Analogue Output (On/Off + Se Calibration check)
Timed Fill On/Off
Timed Fill (Off, 24hr, 48hr, 72hr or select days of week)
Password protection setup
Before proceeding to set-up the data log should be cleared as per section 7.7.
Table 8
Procedure for configuration/calibration the controller
No
Steps of procedure
Display and notes
1
Press
key to enter the menu
The Firmware Version number and serial
number will be briefly displayed followed by the
menu:
2
Press switch 2 to CHANGE SETTINGS
If PASSWORD PROTECTION is disabled
go to step 4
If PASSWORD PROTECTION is enabled the
following message will appear: