Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
34
of
87
No
Steps of procedure
Display and notes
14
Use the + and - keys to set the extra high
alarm delay timer:
the EXTRA HIGH ALARM DELAY timer
setting is adjustable between 0 sec and
120 sec in 5-second steps
Press
key when finished
If the gas bypass timer is set off
(CONFIG/CALIBRATE menu) skip to step
16
If the gas bypass timer is set on (see
CONFIG/CALIBRATE menu) the following
message (example) will appear:
NOTE The gas bypass timer is only used in conjunction with the M507CE-I.
When the refrigerator starts to fill the M507CE-I opens a vent solenoid valve and
discharges gas from the pipe-work. A thermocouple in the outlet senses when liquid
reaches the outlet and switches off the vent solenoid valve allowing the refrigerator to start
filling. If liquid is not detected within a preset time an alarm is sounded and the vent
solenoid valve is closed. This preset time is controlled by the gas bypass timer.
NOTE For further details consult the separate manual (IM1017) or contact Worthington
Industries or your distributor.
15
Use the + and - keys to set the gas bypass
timer:
the GAS BYPASS TIMER setting is
adjustable between 0 min and 30 min in 5-
minute steps
Press
key when finished
The following message (example) will appear:
NOTE The thermocouple in liquid timer is only used in conjunction with the M507CE-I.
When liquid is detected by the thermocouple in the vent outlet the timer is started.
If liquid is still present after the preset time it is possible that the valve is stuck open and an
alarm is sounded.
16
Use the + and - keys to set the
thermocouple in liquid timer:
the T/COUPLE IN LIQUID timer setting is
adjustable between 0 min and 30 min in 5-
minute steps
Press
key when finished
The following message (example) will appear: