Clinispin  HORIZON  products are a  versatile  range  of centrifuges designed with 3 settings to process Chemistry, 
Coagulation or Platelet Poor Plasma and Urine specimens in the same unit. The maximum g-force of 2,000 xg makes 
Clinispin HORIZON compatible with most brands of tubes. Cycle settings can be changed to accommodate custom 
settings.

 

 

F

EATURES 

o

 

The first three (3) cycles are conveniently preset and labelled 

for your lab’s most common applications. Use the 

default cycles or customise them as needed.

 

 

o

 

A time and speed/g-force can be quickly entered for a single use cycle. The cycle will not be retained in memory.

 

o

 

If desired, the control panel can be temporarily locked on one cycle for error-free reproducibility.

 

o

 

A Preset Lock can be turned on to prevent changes from being made accidentally to programmed cycles.

 

o

 

Up to 10 cycles can be programmed for time, speed and braking and labelled with a custom name. Cycles can be 
programmed by g-

force (RCF) rather than speed to facilitate matching validated cycles and tube manufacturers’ 

IFUs.

 

 

o

 

Lid lighting indicates the 

centrifuge’s

 status (ready, running, complete), informing the operator when tubes are 

ready for the analyser and preventing tubes from being left in the centrifuge longer than necessary (patent 
pending).

 

 

o

 

A traditional audible alert indicates the completion of the cycle. 

The audible alert can be muted. 

o

 

Cool

Flow design prevents overheating of samples by using ambient air to keep specimens at room temperature.

 

o

 

The tube holders are fibre reinforced for high strength, durability and years of trouble-free use.

 

o

 

A clear lid permits safe observation of samples and optical calibration of speed.

 

o

 

The lid safety system prevents the centrifuge from operating unless the lid is closed and latched.

 

o

 

The lid safety system only allows entry into the centrifuge after the rotor has completely stopped.

 

o

 

The high-power brushless motor provides years of operation with no routine maintenance.

 

 

I

NTENDED 

U

SE 

General  purpose  laboratory  centrifuge  intended  for  safe  and  rapid  density  based  separation  of  fluids,  including 
physiologic fluids, in approved specimen receptacles for qualitative or quantitative test procedures. As a  general 
purpose  laboratory  centrifuge,  it  is  designed  to  also  run  other  approved  containers  filled  with  chemicals  (non- 
flammable, non-explosive, non-volatile and non-highly reactive only), environmental samples and other non- human 
body samples. This device is intended to be operated by properly trained personnel who have carefully read  the 
operating manual and are familiar with the function of the device. 

 

[Refer to the clinical laboratory method specified by the specimen receptacle manufacturer or established by the 
medical technology for the products applications.] 

 

 

W

ARRANTY 

Woodley Equipment Company warranties that this centrifuge is free from defects in workmanship and parts for 2 
years.

 

Содержание Clinispin HORIZON 12 Flex

Страница 1: ...Clinispin HORIZON 12 Flex 24 Flex Operator s Manual...

Страница 2: ......

Страница 3: ...e without Saving 4 Cycle Lock 5 Single Cycle Lock 5 Preset Lock 5 Setting or Modifying a Saved Cycle 6 Deleting a Cycle 6 Reset to Factory Settings 7 Settings 7 Operation 7 Balancing Loads 8 Care and...

Страница 4: ...dible alert indicates the completion of the cycle The audible alert can bemuted o Cool Flow design prevents overheating of samples by using ambient air to keep specimens at room temperature o The tube...

Страница 5: ...energy and if not installed and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in...

Страница 6: ...or Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycle displayed...

Страница 7: ...op screen displays the speed Setting by G force Press and hold the RCF xg button while changing the displayed setting on the top screen using the up and down buttons next to it The RPM will automatica...

Страница 8: ...and a flash of the lid light will confirm that cycle selection is locked Cancel Single Cycle Lock To re enable cycle selection with lid open press and hold the UNLOCK button for 5 seconds Two beeps an...

Страница 9: ...ons next to the bottom TIME display to adjust time Change Brake Values Navigate with the down key next to the top screen until it shows Brake To change the value shown on the BRAKE display use the up...

Страница 10: ...ATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise to its complete stop position...

Страница 11: ...when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired sample and place it across from this sample Opposing tube holders must be e...

Страница 12: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Страница 13: ...kets have any damage evenslight safely dispose of them and replace them o If the rotor appears damaged contact Technical Support for further assistance Excessive noise or vibration when the centrifuge...

Страница 14: ...If the lid knob still cannot be rotated press the UNLOCK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label a...

Страница 15: ...59 dB A Environmental Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 415 Watts 415 Watts Centrifuge Motor H P Brushless H P Brushless M...

Страница 16: ...their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment a...

Отзывы: