ROUBLESHOOTING 

NOTE: The latch must be turned completely clockwise to its stop position for the centrifuge to operate.

 

 
 

 

The centrifuge does not run

 

 

o

 

Verify that the centrifuge is powered. One of the LED lights 
should be on.

 

o

 

If “Lid not closed” message is displayed, make sure the lid latch 

is turned completely clockwise to its stop position.

 

o

 

If the centrifuge still does not run, contact Technical Support.

 

The rotor does not spin freely 

 

o

 

Make sure nothing has fallen into the rotor chamber, 
following the procedure above.

 

o

 

If nothing obstructs the rotor, the rotor may be damaged. 
Contact Technical Support for further assistance.

 

The centrifuge makes a rattling noise 
when running 

o

 

Stop the centrifuge. Open the lid.

 

o

 

Wearing PPE, remove tubes and tube holders/buckets and 
look for fallen objects or debris. Carefully reach inside the 
rotor chamber with a tool to remove them.

 

o

 

Inspect the rotor, tube holders or buckets for damage.

 

o

 

If the tube holders or buckets have any damage, even slight, 
safely dispose of them and replace them.

 

o

 

If the rotor appears damaged, contact Technical Support 
for further assistance.

 

Excessive noise or vibration when the 
centrifuge is running

 

 

o

 

Verify that all four centrifuge feet are properly seated on a flat 
surface.

 

o

 

Ensure that the load is balanced according to instructions in 

the “Balancing Loads” section of this

 manual.

 

o

 

Make sure that nothing has fallen into the rotor chamber.

 

“Abort” is displayed on the top screen

 

The centrifugation cycle has been interrupted. 

The  centrifuge  stops  and  beeps 
continuously 

The load is not balanced. Press the UNLOCK button, open the lid, 
and balance the load as recommended on page 8 of this manual. 

The centrifuge is stuck on one of the 
settings

 

Cycle selection is locked. Press the UNLOCK button for 5 seconds.

 

Only a few cycles can be accessed

 

The Preset Lock is active. To deactivate it, press the UNLOCK 
button for 5 seconds, until you hear 2 beeps, then again until the 
next 2 beeps. All cycles can now be accessed and/or amended.

 

The cycle time and speed are not set to 
the desired value

 

 

Check the setting by following the instructions in the section on 
Setting or Modifying a Saved Cycle. If the preset is not the desired

 

Содержание Clinispin HORIZON 12 Flex

Страница 1: ...Clinispin HORIZON 12 Flex 24 Flex Operator s Manual...

Страница 2: ......

Страница 3: ...e without Saving 4 Cycle Lock 5 Single Cycle Lock 5 Preset Lock 5 Setting or Modifying a Saved Cycle 6 Deleting a Cycle 6 Reset to Factory Settings 7 Settings 7 Operation 7 Balancing Loads 8 Care and...

Страница 4: ...dible alert indicates the completion of the cycle The audible alert can bemuted o Cool Flow design prevents overheating of samples by using ambient air to keep specimens at room temperature o The tube...

Страница 5: ...energy and if not installed and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in...

Страница 6: ...or Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycle displayed...

Страница 7: ...op screen displays the speed Setting by G force Press and hold the RCF xg button while changing the displayed setting on the top screen using the up and down buttons next to it The RPM will automatica...

Страница 8: ...and a flash of the lid light will confirm that cycle selection is locked Cancel Single Cycle Lock To re enable cycle selection with lid open press and hold the UNLOCK button for 5 seconds Two beeps an...

Страница 9: ...ons next to the bottom TIME display to adjust time Change Brake Values Navigate with the down key next to the top screen until it shows Brake To change the value shown on the BRAKE display use the up...

Страница 10: ...ATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise to its complete stop position...

Страница 11: ...when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired sample and place it across from this sample Opposing tube holders must be e...

Страница 12: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Страница 13: ...kets have any damage evenslight safely dispose of them and replace them o If the rotor appears damaged contact Technical Support for further assistance Excessive noise or vibration when the centrifuge...

Страница 14: ...If the lid knob still cannot be rotated press the UNLOCK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label a...

Страница 15: ...59 dB A Environmental Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 415 Watts 415 Watts Centrifuge Motor H P Brushless H P Brushless M...

Страница 16: ...their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment a...

Отзывы: