adVancEd usEr sETTings (continued):

Braking Rate

 - 

The user may adjust the braking rate from minimum ( slowest deceleration, 0) to 

maximum (quickest deceleration, 9) using the following procedure:

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “BRAKE” is  

 

 

 

displayed in the speed display.

 

3. 

The current braking rate is displayed in the brake display.  Use the UP/DOWN    

 

 

arrow buttons next to the brake display to adjust the braking rate from 0   

 

 

 

(minimum) to 9 (maximum).

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

Countdown Delay

 - 

When the countdown delay is turned on, the timer will not begin 

decrementing until the rotor has reached full speed as set by the user. This feature ensures that the 

samples see full force for the total run time.

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “CNTDLY” is    

 

 

displayed in the speed display.

 

3. 

The current status of this feature (ON or OFF) is displayed in the time display.     

 

 

Press UP/DOWN arrows next to the time display to toggle the status between    

 

 

ON and OFF.

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

imbalance Detection

 - 

This centrifuge is equipped with imbalance detection. This ensures that 

the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction 

occurs.  This feature is intended to provide for operator safety and to extend the life of the 

centrifuge.  The LOW setting will allow for a greater imbalance before shutting the centrifuge down.  

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “BALANC” is    

 

 

displayed in the speed display.

 

3. 

The current status of this feature (LOW or HI) is displayed in the time display.     

 

 

Press UP/DOWN arrows next to the time display to toggle the status between    

 

 

LOW and HI.

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

audible settings

  -  

The user may select from six different audible indicator levels:

Setting  Button Press Beep?  End-of-run audible notification

NO 

NONE

YES 

NONE

NO 

YES, 5 beeps

YES 

YES, 5 beeps

NO 

YES, continuous, user must press STOP/OPEN button to cancel

YES 

YES, continuous, user must press STOP/OPEN button to cancel

 

1. 

While the centrifuge is idle, press the PROGRAM button.

 

2. 

Use the UP/DOWN arrow buttons next to the speed display until “BEEPER” is    

 

 

displayed in the speed display.

 

3. 

The current audible selection is displayed in the time display.  Use the UP/ 

 

 

 

DOWN arrow buttons next to the time display to select a setting (1 to 6).

 

4. 

Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to    

 

 

the speed display to adjust additional settings.

page 11

Содержание 7711039

Страница 1: ...clinispin horizon 853VES Laboratory Centrifuge Rev 1 4 Operator s Manual P N 7711039 ...

Страница 2: ...unctions Time Adjustment and Timer Operation page 10 Speed Adjustment RCF Display page 10 Starting and Stopping a Run page 10 Unlocking the Lid page 10 Advanced User Settings Acceleration Rate page 10 Braking Rate page 11 Countdown Delay page 11 Imbalance Detection page 11 Audible Setting page 11 Memory Locations Storing and Recalling page 12 Digital Cycle Counter page 12 Care and Preventative Mai...

Страница 3: ...d force capability produces platelet poor free plasma and greatly reduces processing times The exclusive Woodley horizontal rotor allows for quick and easy sample loading and complete horizontal separation Save settings in one of 10 memory locations and recall them at the touch of a button Supplied Accessories Standard See next page for optional rotor accessories Also included not shown One 1 10 M...

Страница 4: ...32 100 mm tube holder p n 7713040 75 mm tube holder p n 7713043 250mm Cushion Spacer p n 1525 Part No Max Tube Size Speed Range Force Range 12 Place Performance Plus Horizontal 7786024 Comes with Twelve 12 of each of the tube holders listed 75mm tube holder 7713043 12 17mm x 75mm 500 to 5 000 RPM 40 to 3 100 xg 100mm tube holder 7713040 12 17mm x 100mm 500 to 5 000 RPM 40 to 3 800 xg 12 Place Hori...

Страница 5: ...n RPM Change Buttons RPM Display Running Indicator Light Locked Indicator Light Unlocked Indicator Light Start Button Memory Button Memory Preset Display Program Button Up Down Time Change Buttons Time Display RCF Conversion Button Control Panel see below Cabinet Open Stop Button Brake Display ...

Страница 6: ...e properly grounded 2 For operator safety the locking system is always active requiring power and direction from the user to disengage it the lid also automatically unlocks at the end of a run when it is safe to do so To unlock the lid in order to access the rotor chamber press the OPEN STOP button on the control panel The UNLOCKED indicator light should illuminate If it does not refer to the sect...

Страница 7: ... 11 for a description of these features 7 Begin the run by pressing the START button on the control panel 8 The centrifuge should begin to spin The RUNNING indicator light should illuminate IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 9 After time has elapsed the RUNNING indicator light will extinguish and the rotor will decel...

Страница 8: ...to hold tubes up to 10mL up to 17mm x 100mm The blue holders are de signed to hold tubes up to 5mL up to 17mm x 75mm Your centrifuge must contain a balanced load in order to work properly To ensure that the load is balanced keep these rules in mind when inserting test tube samples into the 12 place horizontal rotor DIRECTIONS 1 Compare the tube to be spun with the three boxes shown below 2 Find th...

Страница 9: ...inserting test tube samples into the 20 place fixed angle rotor 1 Opposing tube holders must be identical 2 Opposing tube holders must be empty or loaded equally weighted samples 3 If an odd number of samples is to be spun use a water filled tube to mate with the unpaired one DIRECTIONS 1 Compare the tube to be spun with the four boxes shown below 2 Find the box that most closely matches the tube ...

Страница 10: ...f the square of the speed this jump size will vary Starting and Stopping a Run With the lid switch closed press the START button to begin a run Press the STOP OPEN button at any time during a run to terminate it Unlocking the Lid The lid is unlocked whenever the red UNLOCKED indicator light is illuminated The lid unlocks automatically for 60 seconds at the end of a run The lid can also be unlocked...

Страница 11: ...s equipped with imbalance detection This ensures that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs This feature is intended to provide for operator safety and to extend the life of the centrifuge The LOW setting will allow for a greater imbalance before shutting the centrifuge down 1 While the centrifuge is idle press the PROGRAM button 2...

Страница 12: ...yed 3 When the desired preset is found pause and the memory display will momentarily flash The stored configuration is now loaded Notes Once a configuration has been loaded it will remain active until any change is made on the control panel If for instance the user adjusts the time the current configuration will no longer match the previously loaded configuration and the memory display will change...

Страница 13: ...aining fragments may provide stress points on subsequent tubes and result in additional breakage If a tube breakage occurs carefully remove the tube holder Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the tube carrier holder Insert a new tube cushion if necessary and replace the tube carrier holder in the rotor 4 Motor and Electrical Mainten...

Страница 14: ...owing table for fault information ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP button to cancel the error and then check the rotor and the line voltage Contact your authorized dealer or Woodley Equipment Company for further assistance ERROR BALANC The centrifuge has detected an imbalan...

Страница 15: ... lid knob enough to cause the pawl to lose contact with the lid safety switch If this happens the centrifuge motor may lose power but the lid will still remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully clockwise to its stop position and the centrifuge will resume operation Circuit Breaker The Horizon is protected with a 4 Amp circuit breaker loc...

Страница 16: ...t parts or accessories 125 mm Tube Holder p n 7713044 Tube Holder Cap p n 7713011 75 mm Tube Holder p n 7713043 100 mm Tube Holder p n 7713040 20 Place Fixed Angle Rotor p n 7786042 50mL tube holder fixed angle p n 7787051 6 Place 50mL Fixed Angle Rotor p n 7786044 250 mm Tube Cushion p n 1525 6mm Tube Cushion p n 9150 Tube Holder Cap p n 7713011 75 mm Tube Holder p n 7713043 100 mm Tube Holder p ...

Отзывы: